13.4 Printing
Symptom
Possible Cause(s) / Solution(s)
Refer To
(Page No.)
Printing Failure
Printing paper may not be installed properly.
→
Open the printer cover and confirm there is no ejected paper
caught inside the printer.
19
Printing paper may be running short.
→
If a red mark appears on the edges of the paper, install a new
printing paper.
19
The batteries may be low.
→
Turn the power OFF and replace the batteries.
6
Printed characters are too light
or faint.
The batteries may be low.
→
Turn the power OFF and replace the batteries.
6
While printing or feeding paper,
the instrument stops working or
restarts.
The batteries may be low.
→
Turn the power OFF and replace the batteries.
→
If you print data often or print out a large amount of data, use
the AC adaptor (sold separately).
23
Only the value shown on the
display is printed out.
Single parameter output mode is activated in the data output
setting.
Switch to all parameters output mode in the function setting.
29
Cannot cancel printing.
You cannot cancel printing.
“E--1” is displayed on the
screen.
The printer paper is not set.
→
Check to ensure printer paper is properly set in the roller.
19
“E--2” is displayed on the
screen.
The battery is low.
→
Turn the power OFF and replace the batteries.
6
Printer Head has overheated due to continuous printer use.
→
Stop printing temporarily and allow printer to cool.
“E--3” is displayed on the
screen.
The Printer Head Lever is up.
→
Pull the lever down.
19
13.5 Digital Output
Symptom
Possible Cause(s) / Solution(s)
Refer To
(Page No.)
Data Transfer Failure
The RS232C cable may not be connected properly.
→
Communication cable exclusively for this model is required.
24
The baud rate setting may not be correct.
Check the settings of the ANEMOMASTER and the computer.
24
Confirm that the communication command is correct.
25
13.6 Analog Output
Symptom
Possible Cause(s) / Solution(s)
Refer To
(Page No.)
Output Failure
Confirm that the polarity of the output terminal is correct.
27
Confirm that normal measurement mode is selected.
11
Air Velocity output has a
staircase pattern.
Check the Time Constant (TC) setting.
13
Incorrect Output Value
Confirm that the analog output setting is correct.
27
Confirm that the output range setting is correct.
27
The load impedance may be set lower than the defined value.
(Load impedance: 5k
Ω
or higher)
27