![Kanomax 6113 Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/kanomax/6113/6113_operation-manual_652686020.webp)
3. Normal Measurement
13
3.4 Changing Time Constant
You can change the Time Constant only when measuring air velocity. The Time Constant for air temperature and
pressure (optional for 6113/6114) is fixed at 1 second.
DISPLAY
PROCEDURE
As you press
key when the Normal Measurement screen (Air
Velocity Measurement Mode) is displayed, the configured time
constant shows up for a moment as shown in the left and then the time
constant for measurement value will be changed.
* The default setting is 1 second.
Every time you press
key, the time constant changes to 1sec,
5sec and 10sec.
You can select Time Constant from 1sec, 5sec and 10sec.
00:01-----Moving average deviation over 1sec will be displayed.
00:05-----Moving average deviation over 5sec will be displayed.
00:10-----Moving average deviation over 10sec will be displayed.
* Once you turn the power off, the time constant will revert to the
default setting of 1 second.
* * * What is Time Constant? * * *
Time Constant is moving average deviations over a certain period of time. If you set the larger value for time
constant, the reading will be stable. On the other hand, if you set the smaller value for time constant, the
reading will be more responsive to the change in air velocity.
This function cannot be used in the Air Temperature Mode and Pressure Measurement Mode.
MODE
How to Take In Measurement Data
EXPLANATION
00:01
(1sec)
Take data 10 times in 1
second
and
display
its
average as an instantaneous
value every 1 second.
00:05
(5sec)
Displays average value over
5 seconds every 1 second.
Data shifts by every 1
second.
00:10
(10sec)
Displays average value over
10 seconds every 1 second.
Data shifts by 1 second.
Time Constant
m/s
0 5 10 15 20sec. (Measuring Time)
Average over 1 sec
0 5 10 15 20sec. (Measuring Time)
Average over 5 sec
0 5 10 15 20sec. (Measuring Time)
Average over 10 sec