3.2 Holding the Reading
DISPLAY
PROCEDURE
While the Normal Measurement screen (Measurement Mode) is on the
screen, press
key.
(You can also use this function in the Air Temperature Measurement
Mode and Pressure Measurement Mode.)
mark will appear at the upper right of the display while the
reading is held.
Press
key again to release the hold mode.
3.3 Changing the Display Range of the Bar Graph
Bar graph shows up only when measuring air velocity.
DISPLAY
PROCEDURE
Press
key when the Normal Measurement mode screen (Air
Velocity Measurement Mode) is displayed.
*
In the default setting the range is
0~10m/s and each scale is equivalent
to 2m/s.
The value on the right end of the bar graph changes. Now the range is
0~25m/s.
(Each scale is equivalent to 5m/s.)
Each time you press
key, the value on the right end of the bar
graph changes. The range changes in the following order:
<Default Setting> 0~10m/s (1scale: 2m/s, 1bar: 0.4m/s)
→
0~25m/s (1scale: 5m/s, 1bar: 1m/s)
→
0~50m/s (1scale: 10m/s, 1bar: 2m/s)
→
0~1m/s (1scale: 0.2m/s, 1bar: 0.04m/s)
→
0~5m/s (1scale: 1m/s, 1bar: 0.2m/s)
→
0~10m/s
* Once you turn the power OFF, the setting will return to the default
setting (0~10m/s).
3. Normal Measurement
12
m/s
m/s
m/s
m/s
m/s
1scale: 2m/s
1bar: 0.4m/s