- 5 -
2. Getting Started
2.1 Power Supply
2.1.1 AC Adapter
Plug the AC adaptor in the inlet. The specification of the provided AC adapter is as
follows (When purchasing AC adapter for yourself, refer to 4. Main Specifications):
Input/Output Specification
:
Input AC 100-240V (50-60Hz)
,
Output DC 12V 0.5A
Plug Specification
:
2.1.2 DC Power Supply
By using No.4 Pin Power (9-28V) and No.5 Pin P-GND of the analog terminal,
DC power can be supplied. (Note: Pay careful attention to the connecting order.)
0.3um
ALM.COM
ALM.ch2
ALM.ch1
P-GND
Power(9-28V)
A-GND
0.5um
1
2
3
4
5
6
7
8
3718-A Analog Terminal
3719-A Analog Terminal
Power also can be supplied through the RS485 connecting terminal. Connecting the
CRMS distributor and RS485 to supply power to the instrument.
For more details on the distributor for the CRMS, please contact your distributor
or your KANOMAX service center.
Inner Diameter 2.1mm
ALM.COM
ALM.ch2
ALM.ch1
P-GND
Power(9-28V)
A-GND
0.5um
1
2
3
4
5
6
7
8
RS-485
DC-IN
5.0um