Component Maintenance Manual
P/N S185X501-XX
25-63-06
Page 7007
© KANNAD 2011
JUN 16/2011
SUBTASK 25-63-06-430-003-A01
(3)
Check of ELT watertightness
CAUTION:
IF AN ELECTRICAL HEATING SOURCE IS USED TO WARM THE BATH, ITS
POWER SUPPLY MUST BE SWITCHED-OFF WHILE HANDLING THE ELT IN
THE BATH.
NOTE:
Not mandatory if O-ring and screws are replaced
(a)
Bring the water of a bath to a temperature of 60 degrees Celsius ± 5 degrees.
(b)
Wholy immerse the ELT in the bath at a depth of 50 cm (19.68 in.).
(c)
Let immersed 5 minutes:
-
check that no bubble rises.
(d)
Remove the ELT from the bath and dry it with compressed air.
SUBTASK 25-63-06-430-004-A01
(4)
Operating tests.
(a)
Perform the ELT Operating Tests procedure described in
)
except for § (6) Current draw
measurement in ARM position
.
NOTE:
DO NOT PERFORM § (6) CURRENT DRAW MEASUREMENT IN ARM
POSITION WHICH REQUIRES A RE-OPENING OF THE ELT.
TASK 25-63-06-400-803-A01
3.
Replacement of Labels
A.
Procedure
CAUTION:
MARKINGS OF ORIGINAL ELTS ARE DIRECTLY ENGRAVED ONTO THE
HOUSING OR FRONT PANEL. THESE MARKINGS CANNOT BE REMOVED.
WHEN PERFORMING A CONTROL, REPLACING A BATTERY, RE-
PROGRAMMING THE ELT FOR THE FIRST TIME, A NEW LABEL HAS TO BE
AFFIXED ONTO THE ORIGINAL MARKING. REMOVING A LABEL IS TO BE DONE
ONLY IF A LABEL HAS ALREADY BEEN AFFIXED DURING A PREVIOUS
CONTROL, BATTERY REPLACEMENT OR RE-PROGRAMMING.
(1)
Check that the new label has all the correct indications and especially the modification index
corresponding to the state of the unit. If necessary, complete the reference number by adding
the modification index.
(2)
Mark the new label with the same serial number as the old one.
(3)
Using tweezers or a scalpel, remove the old label (See CAUTION above), taking care not to
damage the support protection.
(4)
Using a clean, lint-free cloth soaked in alcohol, clean the area where the label is to be fixed so
as to eliminate all traces of grease or dust and to remove the traces of adhesive left by the old
label.
(5)
Remove the paper protecting the back of the new label.
(6)
Affix the new label in the correct position. Make sure that it is in position before applying pressure.
Press lightly to adhere.
The document reference is online, please check the correspondence between the online documentation and the printed version.