10 MAINTENANCE
10.1 CLEANING / DISINFECTING
WARNING: Always disconnect mains power cord before any cleaning procedure.
THE
CONTROL UNIT
Wipe off with a moist cloth and if necessary use your normal cleaning agent.
THE HEATING PAD:
Clean the watertight Heating Pad with soap or your usual cleaning agent. Disinfect the pad with alcohol
or any other disinfectant.
NOTE:
Do not autoclave the pad.
Do not submerge the connector in any liquid.
In case liquid enters the pad connector, the technical department should be
contacted for checking and if required repairing
.
10.2 PERIODICAL INSPECTIONS AND CHECKS
A C T I O N
I N T E R V A L
Inspections carried out by the user/operator.
•
Self test at Power ON
Every time the equipment is
switched on.
•
Visual inspection of pad, cable and connector
Before every use
Checks carried out by technical personnel.
•
Alarm
functions
Once
a
year
Mains failure
Pad failure
Over-temperature
•
Temperature
control
Once
a
year
•
Electrical
safety
Once
a
year
•
Exchange of connector saver
Once a year
10.3 STORING
The Heating Pad shall be stored hanging or lying flat. Avoid folds that may shorten the lifetime.
10.4 TECHNICAL DOCUMENTATION
The supplier will make available on request circuit diagrams, component parts lists, descriptions, or
other information which will assist the user's appropriately qualified technical personnel to exchange
those parts of the equipment which are designated by the manufacturer as exchangeable.
Only the KanMed factory is authorised to perform repairs on the main PCB.