4  DESCRIPTIONS OF FUNCTIONS 

 

4.1  THE CONTROL UNIT 

The Control Unit contains a microprocessor based heating control system that regulates the 
temperature of each of the 8 (4) elements in the Heating Pad. The temperature is continuously 
monitored and the average temperature is displayed in steps of 0.5°C by the temperature LED's. 
 
The safety system in the Control Unit is continuously monitoring the main voltage, internal voltages, run 
parameters, the pad connection, the pad condition and the temperature of each element. 
A safety circuit, totally independent of the microprocessor based operating system, can at any time stop 
the warming should conditions occur that can cause over temperature. 
If a fault is detected the Control Unit will alarm visually and acoustically and switch off the heating. The 
type of alarm is indicated on the alarm display. It may be pad fault, Control Unit fault, over-temperature 
or a combination. 
 

4.2   SELF TEST 

Every time the Operatherm is switched on it will perform a self-test of all functions including the internal 
safety-circuit.  

This is how the self test runs:

 

When pressing the ON /OFF button all front panel indicators and the beeper are activated.  
After about 0.5 sec the red ALARM indicators and the beeper are turned off. 
About 2 sec. later temperature indicators are turned off. 
Simultaneously the red indicators and the beeper are activated again for about 2 seconds. This 
indicates that the safety-test has been successfully completed. 
The Operatherm will then automatically start at 37°C, providing the self test went well. 

 

Do not use the unit if the self test does not perform correctly!

 

 

If a fault is detected during the test the Operatherm will not start automatically and the type of fault be 
indicated by the corresponding alarm LED and the acoustic alarm will sound. 
Switch off the unit and check heating pad connection. Switch on the Operatherm again. If it still alarms 
then switch it off and contact a qualified technician. 

 

 

 

4.3 LAYOUT FRONT 

PANEL

 

 

 
1 Over-temperature 
 
2 Pad 

fault 

 
3 CU. 

Fault 

 

 

Temperature Selection button 

 
5  

Key switch for temp. 

 

>37°C and 32°C 

 
6     To cancel alarm 
 
7     Heating Pad connector 
 
8  

ON/OFF button 

Содержание Operatherm 200

Страница 1: ...ENANCE 12 NOTE Patient warming equipment can harm the patient if not used correctly Never use a unit that do not perform the Self Test according to this manual Make sure that the performance of the unit is regularly checked This manual contains important information and must be kept for future reference It is valid for all Operatherm 202 200 units Manufactured by KanMed AB Gårdsfogdevägen 18B SE 1...

Страница 2: ...it has been examined by a qualified technician The Operatherm 202 is a very safe equipment However certain combinations of heat pressure long term surgery and disinfectant fluids can cause skin damage Therefore it is important to observe the advice regarding pressure relief excessive fluid and temperature selections given later in this section 2 2 EXPLANATION OF SYMBOLS Equipment not suitable for ...

Страница 3: ... safety 4 The Heating Pad cable may not be altered in length or repaired in any other way 5 See Users Manual section 10 concerning cleaning and maintenance of Heating Pads and Control Units 6 Sharp objects must not penetrate the Heating Pad 7 Place the Heating Pad flat on the operating table with the text side towards the patient 8 Gel Pads must be placed over the Heating Pad 9 When used with chil...

Страница 4: ...ent 2 8 TEMPERATURE SETTING Under normal conditions and with KanMed Gel Pads under the patient the maximum temperature 39 C should be selected During long interventions one must consider a lower temperature setting If the OR table surface is hard blood pressure is low i e bad circulation old patients patients with diabetes etc 37 C should be the highest temperature 2 9 USE WITH CHILDREN AND PRE TE...

Страница 5: ... Watertight except connector Surface electrical properties Not electrical conductive Control Unit Voltage 220 240 V AC 100 and110 120 V AC is also available Frequency 50 60 Hz Power 285 VA max Temperature range 32 C and 35 39 C in steps of 0 5 C Accuracy 0 5 C Indicators LED Resolution 0 5 C Over temperature alarm level 40 25 C 0 75 C Dimensions L 255 mm approx W 150 mm approx H 120 mm approx Weig...

Страница 6: ... is switched on it will perform a self test of all functions including the internal safety circuit This is how the self test runs When pressing the ON OFF button all front panel indicators and the beeper are activated After about 0 5 sec the red ALARM indicators and the beeper are turned off About 2 sec later temperature indicators are turned off Simultaneously the red indicators and the beeper ar...

Страница 7: ...long pad or of the 4 elements in the short pad should become warmer than 40 25 C The visual and acoustic alarm will be activated and the over heated element s will be switched off for 90 sec The acoustic alarm can be reset but the over temperature indicator will stay lit If during the 90 sec period the temperature of the alarming element s drops below 40 25 C the visual alarm will automatically ca...

Страница 8: ...ets disconnected during normal use the pad fault indicator and the 32oC LED will be lit The beeper will also sound Switch the unit off and reconnect the Heating Pad Restart the unit and observe a correct self test 4 7 CONTROL UNIT FAULT Any fault in the Control Unit will lit the Control Unit fault symbol and activate the acoustic alarm The heating will be switched off Switch the unit off and try a...

Страница 9: ...y a technician 5 3 SELECTING TEMPERATURE In case you wish to select another temperature than 37 C then do as described in chapter 4 4 1 It will take about 8 15 minutes for the elements in the pad to reach the set temperature depending on surrounding conditions NOTE Please refer to section 2 regarding choice of temperature 5 4 GEL PADS Always use Gel Pads to get an even heat distribution a greater ...

Страница 10: ... 1 2 MAINS POWER FAILURE ALARM AND RESTARTING Connect the Heating Pad to the Control Unit and the Control Unit to the mains Switch on the unit and wait for the self test to be completed Disconnect the power cord The alarm shall sound for at least 10 minutes Reconnect the power cord The unit shall perform the self test and then warm at the temperature that was selected before power off 6 1 3 FAULTS...

Страница 11: ...ntact your local dealer 8 EMC COMPATIBILITY STATEMENT Please see www kanmed se section downloads Operatherm 9 WARRANTY STATEMENT KanMed warrants the purchaser that the Operatherm Control Unit and Heating Pads shall be free from defects in material and workmanship for a period of 12 month from the date of purchase The sole obligation of KanMed with respect to any such defect is limited to the repai...

Страница 12: ...ried out by the user operator Self test at Power ON Every time the equipment is switched on Visual inspection of pad cable and connector Before every use Checks carried out by technical personnel Alarm functions Once a year Mains failure Pad failure Over temperature Temperature control Once a year Electrical safety Once a year Exchange of connector saver Once a year 10 3 STORING The Heating Pad sh...

Отзывы: