Page 33

 

 C.  CALCULATION OF FUEL DATA

 

 
For any fuel not specified by Kane International the net calorific value, gross calorific 
value and composition should be obtained from the fuel supplier. 
 
The following fuel data has been calculated with reference to the efficiency 
calculation. 
 

Example 1:

 

 
Chemical composition: 

 

25% 

 

 

 

 

H

2

   3% 

 

 

 

 

H

2

O  

50% 

 

 

 

 

Q

net

  

8.35 

MJ/kg 

 

 

 

 

Q

g

   9.3 

MJ/kg * 

 

 

 

 

Max CO

2

 20.4% 

 

K1n

  = (255 x % carbon in fuel) / Q

net

 (kJ/Kg) 

 

= (255 x 25) / 8350  =  

0.763

 

 
 

K1g

  = (255 x % carbon in fuel) / Q

g

 (kJ/Kg) 

 

= (255 x 25) / 9300  =  

0.685

 

 
 

K2

 

= Max % CO

2

 =  

20.40

 

 

K3

 

= Wet Loss 

=  [(9 x %H

2

 + %H

2

O) / 9300] x 2425 

 

 

   

=  [(9 x 3 + 50) / 9300] x 2425 

 

 

   

=  (77 / 9300) x 2425  =  

20.08

 

 

K4

 = 

65

 (an approximation for wood)  * 

 
The fuel values to program into the Analyser are as follows: 
 

 

NATURAL GAS 

K1g 
K_2 
K_4

0.763  

:

 20.4 

:

   65 

K1n 
K_3 
O2r 

:

 0.685 

:

 20.08 

:

  8.0 

 
 

 

* Assumed values in the absence of supplied data. See previous appendix for other 
fuels. 
 
 

Keison Products     www.keison.co.uk       [email protected]      Tel: 01245 600560      Fax: 01245 600030

Содержание KANE900 Plus

Страница 1: ...Kane House Swallowfield Welwyn Garden City Hertfordshire AL7 1JG Tel 44 0 1707 375550 Fax 44 0 1707 393277 E mail sales kane co uk Stock No 18310 March 2006 Kane International Limited Keison Products...

Страница 2: ...0 4 5 Taking a Pressure Reading 10 4 6 Regular Checks During Sampling 11 4 7 Normal Shutdown Sequence 11 4 8 Electromagnetic Compatibility 12 5 MOVING THROUGH THE MENUS 13 19 5 1 Basic Operation 13 5...

Страница 3: ...2 Changing the particle filter 26 10 PROBLEM SOLVING 27 11 CALIBRATION 27 12 PRODUCT SPECIFICATION 28 APPENDICES A MAIN DISPLAY PARAMETERS 29 B COMBUSTION EFFICIENCY CALCULATION 31 C CALCULATION OF FU...

Страница 4: ...wn through options PUMP Turns pump on and off STORE Enters data storage menu ENTER Accepts a command ie enters a menu option PRINT Prints current data RS232 8 pin din Charger socket Store Print Pump E...

Страница 5: ...KANE900 Plus Operators Manual Page 2 1 2 Instrument Layout Rear Keison Products www keison co uk sales keison co uk Tel 01245 600560 Fax 01245 600030...

Страница 6: ...KANE900 Plus Operators Manual Page 3 1 3 Standard Probe Configuration Keison Products www keison co uk sales keison co uk Tel 01245 600560 Fax 01245 600030...

Страница 7: ...ual Page 4 1 4 Analyser Connections Gas Connection to Instrument Particle Filter In line Water Trap Neoprene Hose Flue Temperature Connector Keison Products www keison co uk sales keison co uk Tel 012...

Страница 8: ...p to 31 C decreasing linearly to 50 RH at 40 C Mains supply fluctuations not to exceed 10 of the nominal voltage 3 FIRST TIME USE Charge the battery for 12 hours Following this an overnight charge sho...

Страница 9: ...be set to zero and the Oxygen sensor to be set to 20 9 After switch on the analyser will briefly display header information Kane International 44 1707 375550 And then show the countdown screen ZERO CA...

Страница 10: ...Press This zeros the toxic sensor and sets Oxygen to 20 9 The next screen is the MAIN DISPLAY of the analyser NETT C 0 0 O2 20 9 CO ppm 0000 EFF G 0 0 Use and to change the display CO2 0 0 FLUE C 0 0...

Страница 11: ...between the different modes is detailed in Display Menu Section 5 2 4 4 3 1 4 Page Mode Use the and keys to change the information that is displayed on the screen The following pages are available NAT...

Страница 12: ...move parameter up O2 20 9 CO ppm 0000 CO2 0 0 CO2 0 0 Select next parameter Repeat above until display reads desired data O2 20 9 CO ppm 0000 CO2 0 0 CO CO2 R 0 0001 4 3 3 8 Page Mode Displays 8 param...

Страница 13: ...be used in holes whose diameters range from 8 mm to 21 mm 5 16 to 13 16 inch The standard probe is rated at 650 C 1202o F Temperatures of up to 1100 C 2012 F can be accommodated using an optional high...

Страница 14: ...be is moved Take care and watch the water trap closely The in line particle filter is clean and does not become blocked 4 7 Normal Shutdown Sequence DO THIS EVERY TIME YOU USE THE ANALYSER Remove the...

Страница 15: ...se The following procedure should be adopted Go through the normal start up sequence in the location where the equipment is to be used Switch on all localised electrical equipment that might be capabl...

Страница 16: ...ted Menu MAIN MENU 1 SELECT 3 DISPLAY 2 UNITS 4 SETUP Press to select parameter FUEL O2 Ref SMOKE RESET LIGHT OIL OFF OFF NO Use and to change setting i e fuel selected FUEL O2 Ref SMOKE RESET NATURAL...

Страница 17: ...2 2 Select Menu FUEL O2 Ref SMOKE RESET NATURAL GAS OFF OFF NO This menu allows selections to be made for the parameters detailed below FUEL Select the fuel being used by the boiler from either a sta...

Страница 18: ...then toxic gas measurements will be displayed with the symbol n attached to the reading i e CO n What does Oxygen reference mean If 3 O2 reference is selected and 5 O2 is measured in the flue then to...

Страница 19: ...anges the toxic gas measurement units Select from volumetric readings parts per million ppm or mass flow reading milligrams per cubic meter mg m3 PRESS Flue draught can be displayed in millibar mbar h...

Страница 20: ...tion time CO gas alarm NOx percentage for calculation Date and time Printout Header LANG CO AL CALEN ENGLISH ARM 400 DAR ZERO 3 NOx 5 HEADER Parameter Description Settings LANG Changes the analysers d...

Страница 21: ...l be shown once the parameter is entered TIME FORMAT DATE hh 13 dd 03 mm ss 53 26 mm yy 01 96 FORMAT Changes the date format for display and printing dd mm yy yy mm dd mm dd yy To change the time posi...

Страница 22: ...STORE KEY The screen above shows the standard header setting with the cursor now shown underlining the K in Kane By using and any letter or number can be chosen Once the correct character is displaye...

Страница 23: ...ta may be lost if transmission is interupted Keep the KANE900 Plus pointing at the printer until the printout has finished Dot matrix serial printer requires the supplied serial cable to transmit data...

Страница 24: ...KANE900 Plus Operators Manual Page 21 6 2 Standard Printout The standard printout is Keison Products www keison co uk sales keison co uk Tel 01245 600560 Fax 01245 600030...

Страница 25: ...the next test On the display shown above the next location will be 3 To store a test set MODE to STORE and press The current readings will be stored in the analysers memory Tip Make a note of the loc...

Страница 26: ...shed and the display below is shown Press to exit the print routine The instrument will return to main display PRINT TESTS 1 to 10 PLEASE WAIT MENU TO ESCAPE 7 3 Deleting Data To delete the data in st...

Страница 27: ...ng you can press when you are ready to store the next sample In addition the is now a This is visible even if the countdown line is not and is there to indicate that the instrument is ready to accept...

Страница 28: ...he average of three viewing mode Ave View as above Offer a printer to the instrument in the usual way Press The print out omits any reading flagged as NOT FITTED Keison Products www keison co uk sales...

Страница 29: ...ills onto the skin or clothing clean off immediately using fresh water seek medical advice if problems occur 9 2 Changing the Particle Filter This is a very important part of the analyser and should b...

Страница 30: ...pond to flue gas Particle filter blocked Probe or tubing blocked Pump not working or damaged with contaminents Probe connected to pressure connector Flue temperature readings erratic Temperature plug...

Страница 31: ...m 5 full scale 5 0 full scale 0 3 reading 1 0 reading 1 0 reading 0 2 2 C 0 3 reading 0 0001 0 01 0 21 0 4000ppm 0 5000mg m3 0 10 000ppm 0 12 000mg m3 0 5000ppm 0 6700mg m3 100ppm 200mg m3 150 mbar 0...

Страница 32: ...ELECT menu 5 2 2 for oxygen reference The display will read O2 20 if referenced values selected and instrument is in clean ambient air EFF G Combustion Efficiency calculation displayed in percentage G...

Страница 33: ...t O2 20 will be displayed while in clean ambient air Prs Flue draught pressure reading Displayed when pressure sensor fitted See UNITS menu 5 2 3 for scales NO Nitric Oxide reading in ppm or mg m3 Dis...

Страница 34: ...iciency calculations assume that the energy contained in the water vapour is not recovered Since the fuel air mixture is never consistent there is the possibility of unburned partially unburned fuel p...

Страница 35: ...ry flue gas losses wet losses 100 20 9 x K1g x Tnet K2 x 20 9 O2m K3 x 1 0 001 x Tnett Excess Air 20 9 20 9 02m 1 x 100 CO2 20 9 O2m x K2 20 9 Unburned fuel Loss K4 x CO CO CO2 Note CO scaled in Where...

Страница 36: ...MJ kg Qg 9 3 MJ kg Max CO2 20 4 K1n 255 x carbon in fuel Qnet kJ Kg 255 x 25 8350 0 763 K1g 255 x carbon in fuel Qg kJ Kg 255 x 25 9300 0 685 K2 Max CO2 20 40 K3 Wet Loss 9 x H2 H2O 9300 x 2425 9 x 3...

Страница 37: ...3 EN 61000 6 1 and is certified to be compliant Specification EC EMC KI KANE900 Plus details the specific test configuration performance and conditions of use Please Note Batteries used in this instr...

Страница 38: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: