REGULAR CHECKS DURING SAMPLING
Do not exceed analy
z
er operating specifications:
•
Do not exceed probe maximum temperature
•
Do not exceed analy
z
er internal temperature
•
Do not place analyser on a hot surface
•
Keep analy
z
er water trap vertical
-
water vapour condenses in probe line
&
can quickly fill analyser water trap
•
Keep analy
z
er in
-
line particle filter clean
&
dry
DO THIS EVERY TIME YOU USE THE ANALY
Z
ER
Remove probe from exhaust
-
TAKE CARE! HOT PROBE
-
&
allow to cool naturally.
Allow analy
z
er to purge in fresh air for at least three minutes or until all
toxic sensor readings are below 10ppm.
Do not immerse probe in water as this will damage pump
&
sensors.
NOTE:
It is good practice to hang probe hose vertically after sampling so
condensate drains away.
KANE
-
EGA4/5 MANUAL 29
12
13
NORMAL SHUTDOWN SEQUENCE