ER
CHECK BEFORE SWITCH
-
ON:
Particle
&
water stop filters are dry and clean
Water trap
&
probe line are empty of water
Water trap is correctly fitted
&
instrument upright
All hoses connections, etc, are properly made
Analyser & probe will sample fresh outdoor air
Power ON by pressing
to start automatic calibration count down.
AUTOMATIC CALIBRATION
During automatic calibration analy
z
er samples fresh air to zero toxic
sensors
&
set oxygen sensor to 20.95%.
After power on your analyser displays identity, software version
&
serial
number.
“
ANALY
Z
ER PURGING 90 secs
”
countdown appears on display.
Calibration time counts down in seconds to zero
&
can be changed to 90,
120, 180 or 300 seconds.
NOTE:
-
180 seconds is recommended to allow sensors to fully stabili
z
e
-
anything less may result in toxic
&
oxygen sensor drift.