10
Nom. flow
Q
n
test point
Q
n
/Q
min
/Q
t
[m³/h]
test quantity
[kg]
test time
[min]
0.6 m³/h
0.6 / 0.006 / 0.036
30 / 2 / 3.6
3 / 20 / 6
1.5 m³/h
1.5 / 0.015 / 0.09
75 / 5 / 10
3 / 20 / 6.7
2.5 m³/h
2.5 / 0.025 / 0.015
125 / 10 / 20
3 / 24 / 6
3.5 m³/h
3.5 / 0.035 / 0.21
175 / 10 / 21
3 / 17 / 6
6.0 m³/h
6.0 / 0.06 /0.36
300 / 20 / 40
3 / 20 / 6.7
10 m³/h
10 / 0.1 / 0.6
500 / 20 / 60
3 / 12 / 6
15 m³/h
15 / 0.3 / 1.5
750 / 50 / 150
3 / 10 / 6
25 m³/h
25 / 0,5 / 2,5
1250 / 50 / 250
3 / 6 / 6
40 m³/h
40 / 0.8 / 4
2000 / 100 / 400
3 / 7.5 / 6
60 m³/h
60 / 1.2 / 6
3000 / 200 / 600
3 / 10 / 6
150 m³/h
150 / 3 / 15
7500 / 500 / 1500
3 / 10 / 6
400 m³/h
400 / 8 / 40
20,000 / 1250 / 4000
3 / 9.4 / 6
1000 m³/h
1000 / 20 / 100
50,000 / 2000 / 10,000
3 / 6 / 6
Test parametres are set up in accordance
with German regulations relating to accur-
acy class C.
The individual test time is selected on the
basis of:
Min test time of 3 min (Q
n
)
Water quanitity of at least 10% of water
quantity pr hour (Q
t
and Q
min
)
Water quantity for Q
t
- corresponding to
min. 1000 pulses
Water quanity for Q
min
corresponding to
min 500 pulses
These suggestions can be optimised for
each individual bench and, of course, with
consideration for the test objectives.