Once RealTerm has been installed and configured as per the instructions on the "RealTerm Software Configuration"
section, any of the commands listed in the table below can be used to view or change the various instrument's
parameters.
COMMAND LIST
COMMAND
PURPOSE
^ostart
Start continuous output
output
Single line output
^ostop
Stop continuous output
lcal (lower case L)
Live calibration
^pcal,<value>,<value2>
Calibration for laboratory sample analysis
?ppv
View the process parameters values
=degc
Set the temperature units displayed to Celsius (°C)
=degf
Set the temperature units displayed to Fahrenheit (°F)
^pct20, <value>
Set the percent water for the 20 mA output
^alarm, <value>
Set high water percent alarm value
^alhy,<value>
Set alarm hysteresis (lower limit for the alarm activation band)
=alnc
Set alarm mode to normally closed
=alno
Set alarm mode to normally open
=date,<MM>/<DD>/<YYYY>
Set the current date
=time,<HR>:<MM>::<SS>
Set the current time
^modbus
View/Change Modbus parameters
?sn
View the serial number of the instrument
?version
Display the software version of the instrument
help
Display a list of commands
The command the symbol "^" can be replaced with the following options:
"*" to see the command usage details
"?" to see the current value of the parameter
"=" to change the value of the parameter
Type the command with the replaced symbol in the command box and click on "Send ASCII" to send the command.
Detailed instructions on how to use commands and examples are available in the previous sections.