Full Digital Mechanical Sector Scanning Ultrasonic Diagnostic Instruments (Vet) User’s Manual V2.14
- 6 -
5.2.1 Ultrasonic probe connection
Insert the probe connector into the probe socket in the back of the main unit;
5.2.2 Ultrasonic probe disconnection
Socket screwdriver
Button position
on both side of socket
According to arrow direction,
gently insert the socket
screwdriver, then button up the
button position on both side of the
socket with your fingers, you can
remove the probe socket.
Fig. Remove the probe
5.3 Install/Remove the battery
1. Install the battery
Place the main unit on the plane, according to the direction shown (finger upward direction, the plate
of battery toward right) put the battery into battery storage of main unit, cover the “Battery cover”
into the corresponding position, and put “Battery lock coin” into the groove of the “Battery cover
knob”, and then toggle the “Battery cover knob” according to “CLOSE” direction of arrow to lock
the “Battery cover”.
Warning:
Before connecting or using the probe, make sure that the probe, connecting
cable and connector are in normal condition (free of cracks or drop). Use of defective
probe may cause electric shock.
Turn off the system, and use socket
screwdriver to remove the probe with
figure’s method and instructions.
Warning:
1. The Kaixin ultrasonic probe shall be connected to the dedicated Kaixin ultrasonic system
only. Select proper probe model according to the relevant instructions of ultrasonic
diagnostic system.
2. Check the ultrasonic probe and connecting cable after diagnostic operation. Use of
defective probe may cause electric shock.
3. Do not knock or bump the probe, or it may be damaged and cause electric shock.
4. Unauthorized disassembly of the probe shall be prohibited as it may cause electric shock.
Attention:
1. Turn off the ultrasonic system before disconnecting the probe. Disconnecting the probe
with system power on may damage the system or probe.
2. Before disconnecting the probe, place the cable and probe on a stable and leveled position
so that the probe may not be damaged or injury person by unexpected fall.
3. Freeze the instrument when instrument is start-up without operation to increase of service
life of probe.
4. Repeat available machine time should be more than 5 minutes to avoid turn on/off power
supply in short time.
Danger:
Use together with flammable anaesthetic, it may result in explosion.
Attention:
Probe is highly sensitive to shake, be used with caution. About probe’s use
and cleaning, the details see the relevant sections.