Page 22
KAHN INSTRUMENTS, INC.
885 Wells Road, Wethersfield, CT 06109
Tel: 860-529-8643; Fax: 860-529-1895
E-mail: [email protected]
www.kahn.com
4.5
User Inputs / Outputs
4.5.1 Analog Inputs / Outputs
The monitor unit features, as standard, the following inputs and outputs. These are
available on the “User I/O” connector located at the rear of the monitor unit.
Pin
Number
Function
Range
Notes
1
0V Analog
2
Dewpoint 4-20mA
-60 to +30
°
Cdp
Max 500
Ω
Load
3
Dewpoint 10mV/
°
C
-100 to +100
°
Cdp
4
Temperature 4-20mA
-50 to +50 °Cdp
Max 500
Ω
Load
5
Temperature 10mV/
°
C
-80 to +80
°
C
6
% RH 4-20mA
0 to 100 %
Max 500
Ω
Load
7
% RH 0-1 V D.C.
0 to 100 %
8
0V Digital
9
Data Hold TTL
5V = Data Held
Digital Output
10
ABC TTL
5V = Balancing
Digital Output
11
Optics Alarm
0V = Alarm
Digital Output
12
Remote ABC Initiate
0V to initiate
Input
The “USER I/O” connector located on the Monitor rear panel is a 12 way screw-terminal
type and accepts wire pin or bare wire connections.