2.13 Pressure Measurement
A remote pressure transducer may be connected to the S4000
Integrale
Monitor. This
allows on-line measurement and display of the sample gas pressure, as well as
calculation of pressure compensated PPM(V), PPM(W), g/m
3
, and g/kg.
If a pressure transducer is not connected to the instrument, a default pressure of 101.3
kPa (atmospheric pressure) is assumed, and all pressure dependent parameters are
calculated using this pressure.
The pressure is displayed in kPa, and is available as a digital output on the RS232
interface. Refer to Section 4.5.2.
The standard pressure range is 0 to 50 psi absolute, but special pressure ranges may be
supplied to order. Contact Kahn Instruments Technical Sales Department for further
information.
Remote Pressure Connection
A 6 pin circular socket (type Hirose RM12B Series bayonet) on the monitor unit rear
panel marked REMOTE PRESSURE provides the supply voltage (+15VDC) and signal
input voltage (1-6VDC) scaled over the range of 0-50 psia required for the pressure
measurement circuit.
Pin connections are as follows:
Pin Number
Function
1
N/C
2
N/C
3
N/C
4
Signal Input
5
0 VDC
6
+15 VDC