OPERATION
4.4.4 STANDBY Mode
In STANDBY mode, the Peltier heat pump is disabled.
The main use for this feature is during set up (when measurements are not required),
i.e. when flow rates are being adjusted and the analog outputs are being configured.
4.4.5 Parameter Conversions & Pressure Compensation
Many parameters which are calculated by the Optidew require a temperature or pressure
reading in addition to dewpoint to ensure the calculated value is correct.
These additional readings can either come from a sensor connected to the Optidew, or
from a fixed (manual) input. See section 3.2 for details on external inputs.
Calculated Parameter
Temperature input
required
Pressure input required
%RH
wvp
g/m
3
g/kg
Wetbulb
ppm
V
ppm
W
If external sensors are used to generate the inputs, then the sensors should be
positioned in the same location so that they are making a measurement representative
of the environment seen by the dewpoint sensor.
4.4.6 Data Logging
The data logging function allows all of the measured parameters to be logged at a
user specified interval on the supplied SD card via the SD card slot on the base or side
of the instrument. The filename for each log file is generated automatically from the
instrument date and time.
Log files are saved in CSV (comma separated value) format. This allows them to be
imported easily into Excel or other programs for charting and trend analysis. To set-up
data logging refer to Section 3.2.
4.5
Minimum Measurable Dewpoints
The minimum dewpoint that can be measured is determined by the sensor temperature,
and whether the sensor can be maintained at that temperature. The following chart
assumes operation in a climatic chamber, where the air speed is sufficient to remove
any excess heat generated by the sensor.
Содержание Optidew Series
Страница 2: ...Optidew For Kahn Instruments contact information please go to www kahn com...
Страница 13: ...Optidew User s Manual 4 October 2019 INSTALLATION This page is intentially left blank...
Страница 53: ...Optidew User s Manual 44 October 2019 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...Optidew User s Manual Kahn Instruments 47 APPENDIX B Appendix B Modbus Register Map...
Страница 65: ...Optidew User s Manual 56 October 2019 APPENDIX C Appendix C Dimensional Drawings...
Страница 70: ...Optidew User s Manual Kahn Instruments 61 APPENDIX C Dual Stage Dewpoint Sensor...
Страница 71: ...Optidew User s Manual 62 October 2019 APPENDIX C Sample Block...
Страница 74: ...Optidew User s Manual Kahn Instruments 65 APPENDIX C...
Страница 75: ...Optidew User s Manual 66 October 2019 APPENDIX C...
Страница 76: ...Optidew User s Manual Kahn Instruments 67 APPENDIX C...
Страница 77: ...Optidew User s Manual 68 October 2019 APPENDIX C...
Страница 78: ...Optidew User s Manual Kahn Instruments 69 APPENDIX C...
Страница 79: ...Optidew User s Manual 70 October 2019 APPENDIX C...
Страница 80: ...Optidew User s Manual Kahn Instruments 71 APPENDIX C...
Страница 81: ...Optidew User s Manual 72 October 2019 APPENDIX C...
Страница 82: ...Optidew User s Manual Kahn Instruments 73 APPENDIX C...
Страница 83: ...Optidew User s Manual 74 October 2019 APPENDIX C...
Страница 84: ...Optidew User s Manual Kahn Instruments 75 APPENDIX C...
Страница 85: ...Optidew User s Manual 76 October 2019 APPENDIX C...
Страница 86: ...Optidew User s Manual Kahn Instruments 77 APPENDIX C...
Страница 87: ...Optidew User s Manual 78 October 2019 APPENDIX C...
Страница 88: ...Optidew User s Manual Kahn Instruments 79 APPENDIX C...