Easidew Transmitter User’s Manual
Kahn Instruments
5
INSTALLATION
2.4.1 Electrical Boundaries
100
200
300
400
500
600
12
14
16
18
20
22
24
26
28
Resistanc
e (ohms)
Supply Voltage
Figure 6
Maximum Load of Easidew - Including Cable Resistance
2.5
Transmitter Mounting
Prior to installation of the transmitter, unscrew and remove the blue plastic cover and
retain for future use. Take care to prevent any contamination of the sensor before
installation (handle the transmitter by the main body only, avoiding contact with the
sensor guard).
The Easidew can be mounted into either a flow-through sensor sampling block (optional)
or directly inserted into a pipe or duct and can be operated at pressures of up to 5000
psig when installed with the bonded seal provided.
The recommended gas flow rate, when mounted in the optional sampling block, is 2 to
10 scfh (1 to 5 Nl/min). However, for direct insertion applications, gas flow can be from
static to 33 fps (10 m/sec).
NOTE: Place the bonded seal over the 5/8”- 18 UNF mounting thread and
assemble into the sampling location by hand using the wrench flats only. DO
NOT grip and twist the sensor cover when installing the sensor.
When installed, fully tighten using a wrench until the seal is fully compressed and to the
following torque setting:
5/8” - 18 UNF
22.5 ft-lbs (30.5 Nm)