AV1 User Manual
3
1
Unpacking and inspection
Unpack the equipment.
Check the packing and picking list to insure all accessories are there.
Examine the packing material carefully for separately packed items.
Check the table for external surface damage.
Please be aware that the springs must not be touched at any time. Please use gloves
when you assemble the unit.
2
General description
The K-SYSTEMS Anti Vibration tables are designed to meet the requirements of ICSI
Procedures. With an isolated microscope platform and a complex anti-vibration
system, the table is preventing vibrations from reaching the microscope platform and
dampening unavoidable vibrations in the shortest possible time.
The inverted microscope is placed on an individually suspended floating platform.
The platform must be levelled and isolated – meaning not touching – the rest of the
table plate. The vibrations are reduced by friction between the platform and the
springs holding the platform.