36
Operation
Table 3-2: Analog Input Range and Output Digital Code (cont’d)
3.5 ADC Sampling Rate
Sampling Rate refers to ADC internal conversion speed as set by the user.
When programming through a software API, the desired ADC sampling
rate must be set, whether for single value, using a software polling com
-
mand, or block data in continuous buffer mode. Available sampling rates
are 20SPS, 40SPS, 80SPS, 160SPS, 320SPS, 500SPS, 1000SPS, and
2000SPS.
3.5.1
Software Polling Data Transfer (Non-Buffering Pro
-
grammed I/O)
Polling mode benefits flexible timing and is suitable for retrieving the
latest data without FIFO buffering latency. The USB-62401 continuously
updates the latest acquired data onto a data port for specific channels.
Data not retrieved in time is overwritten with new data without notice.
As the software polling rate (here equaling data rate) of a PC may exceed
the ADC sampling rate, it is possible to receive multiple identical data
before a new conversion has completed. Please refer to UD-DASK func
-
tion reference for the details of corresponding software API instruction.
Description
Bipolar Analog Input
Range
Digital Code
Full-scale range
±78.125 mV
N/A
Least significant bit 9.313nV
N/A
FSR-1LSB
78.1249907 mV
7FFFFF
Midscale
+1LSB
9.313nV
000001
Midscale
0 V
000000
Midscale
-1LSB
-9.313nV
FFFFFF
-FSR
-78.125mV
800000
NOTE:
NOTE:
Accuracy frequently deteriorates with increased ADC sampling rate.
Содержание USB-62401
Страница 2: ...ii...
Страница 8: ...viii List of Figures This page intentionally left blank...
Страница 10: ...x List of Tables This page intentionally left blank...
Страница 21: ...Introduction 11 Figure 1 3 USB 62401 Module Side View...
Страница 22: ...12 Introduction Figure 1 4 USB 62401 Module Front View...
Страница 24: ...14 Introduction Figure 1 7 Module In Stand Front View...
Страница 25: ...Introduction 15 Figure 1 8 Module Stand Top View 20 4 20 4 B 26...
Страница 29: ...Introduction 19...
Страница 30: ...20 Introduction This page intentionally left blank...
Страница 34: ...24 Getting Started Figure 2 4 Module Pre Rail Mounting Figure 2 5 Module Rail Mounted...
Страница 36: ...26 Getting Started This page intentionally left blank...
Страница 43: ...Operation 33 Figure 3 7 4 Wire RTD Connection Figure 3 8 3 Wire RTD Connection...
Страница 56: ...46 Calibration This page intentionally left blank...
Страница 62: ...52 Getting Service...