PCIe/PXIe-6301 Series |
| 28
Conversion of Resistance and Temperature
The PCIe/PXIe-6301 is a temperature measurement module designed for the PT100
RTD
with a maximum measurable resistance of 400 Ω. During temperature
measurement, PCIe/PXIe-6301 first measures the resistance value, then automatically
converts the resistance
value (Ω) to the temperature value (°C)
.
For the conversion of resistance and temperature values, the internal conversion
formula of the driver is referenced to the platinum RTD standard DIN/IEC 60751:2008
(Edition 2.0), ie:
When the temperature is between -200 °C and 0 °C:
R
t
= R
0
[1 + At + Bt
2
+ C(t − 100
℃
)t
3
]
When the temperature is between 0 °C and 850 °C:
R
t
= R
0
(1 + At + Bt
2
)
Where:
⚫
R
t
: The resistance value (
Ω
) of PT100 when the temperature is t (°C)
⚫
R
0
: The resistance value (
Ω
) of PT100 when the temperature is 0 °C, ie. 100
Ω
⚫
A = 3.9083 ∗ 10
−3
℃
−1
⚫
B = −5.775 ∗ 10
−7
℃
−2
⚫
C = −4.183 ∗ 10
−12
℃
−4
The complete temperature/resistance index table is shown in
Get resistance value
The resistance value
(Ω)
can be read from the PCIe/PXIe-6301 when the ReadRawData
method is called.