12.2 Error messages
All error messages are displayed flashing
on the matrix display at regular intervals.
The other instrument functions remain unaf-
fected as far as posssible.
Display
Cause / Remedy
Status message
NO PAPER!
The chart cassette has been removed or the end of the
chart has been reached; a fresh chart has to be fitted.
Section 9.2 ff
Caution: Insert the chart cassette by applying light up-
wards pressure
The take-up roll has not properly engaged with the
chart cassette
Section 9.2.1
Error or fault on recorder
LOW BATTERY!
The battery for the real time clock and for RAM back-
up is discharged.
Please contact the nearest offfice or the factory.
ADJUST TIME!
This errror message is produced when the storage ca-
pacitor (Code 021,
Section 2.2) is discharged too far
during supply failure.
Event, supply failure and operating hour counters and
the system clock are reset.
The time must be set.
Section 8.3.2
RELAY-ERROR!
The communication with the external relay module
ER8 is faulty.
Check the connection between the recorder and the
external relay module ER8.
Содержание LOGOLINE 500d
Страница 106: ...9 Consumables 106 9 2 Removing and replacing the chart cassette S F ENTER E S F ENTER E...
Страница 108: ...9 Consumables 108 9 2 1 Changing the roll chart 1 5 2 6 3 7 4 8...
Страница 109: ...9 Consumables 109 9 2 2 Changing the fanfold chart 1 4 2 5 3...
Страница 110: ...110 10 Extra Codes and accessories 10 1 Converting the chart cassette 1 3 2 4...
Страница 113: ...10 Extra Codes and accessories 113 ER 8 Error Power L L L1 N 97 98 99 TxD RxD RxD TxD GND 8 1 8 2 8 3...
Страница 136: ......
Страница 137: ......
Страница 138: ......
Страница 139: ......