11 Fault finding
120
Relay does not switch
although limit is exceeded
-
The signal inputs are not wired up according to the connection
diagram
-
Pen recorder and external relay module ER8 are not connected
together correctly
-
The switching differential at the limit has been disregarded
-
The relay status is “OFF”
-
A wrong signal is being monitored against the limit
No programming possible
-
A wrong codenumber has been input
-
The codenumber has been altered
No communication with the
setup program
-
The setup connector has not been plugged in correctly
-
The recorder is already being configured using the keys; leave
parameter and configuration level by pressing the
e
key.