
11 Calibrating a sensor with a standard signal
84
The transmitter is in "measuring mode" and displays the compensated
conductivity of the solution.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
11.5 Concentration operating mode
11.5.1 Calibration of the relative cell constant
The transmitter is in "Measuring mode".
✱
Immerse the conductivity sensor in a sample medium with a known
conductivity.
✱
Start the calibration, see "Ways to start the calibration", page 71.
✱
Press the
key.
The measured conductivity value is displayed.
✱
Wait until the measurement value has stabilized.
This example is based on a conductivity sensor with a connected two-wire
transmitter.
The conductivity of a caustic solution is converted into a concentration value
[%] by the device.
PGM
Содержание dTRANS pH 02
Страница 66: ...9 Calibrating a redox measurement chain 66...
Страница 89: ...89 12 Calibration logbook Next most recent successful calibration Briefly press the key...
Страница 100: ...14 Setup program 100 Export data for processing in an external program...
Страница 133: ......
Страница 134: ......
Страница 135: ......