11 Calibrating a sensor with a standard signal
✱
Enter the required end temperature (2) for the temp. coef. curve.
✱
Heat the sample medium continuously
(3) the current uncompensated conductivity
(4) the current temperature of the sample medium
(5) the first target temperature
During the calibration process, the device displays values for the following five
temperature interpolation points.
The end temperature has been reached
Use the
key to accept the temperature coefficients or
the
key to reject the calibration result.
The LC display now shows the determined temperature coefficients as %/K.
✱
Use the
key to accept the temperature coefficients or
the
key to reject the values.
(2)
(3)
(4)
(5)
During calibration, the rate of temperature change in the measurement
solution must not exceed 10 °C/min.
PGM
EXIT
PGM
EXIT