5 T
echnical data
17
5
Technical data
Meas. input
Designation
Meas. range
Meas. accuracy
1
/
ambient
temperature error
Recognition of ...
Probe short-
circuit
Probe break
Resistance
thermometer
Pt100 EN
60 751
-200 to +600°C
0.1%/
≤
100ppm/°C
recognized
recognized
Pt1000 EN 60 751
-200 to +600°C
0.1%/
≤
100ppm/°C
recognized
recognized
KTY2X-6 (PTC)
-50 to +150 °C
1%/
≤
100ppm/°C
recognized
recognized
resistance 0 — 3000
Ω
customer table
3
0.1%/
≤
100ppm/°C
3
= 0
Ω
recognized
Measuring current with Pt100: 0.2 mA, with Pt1000, KTY2X-6 or resistance: 0.02 mA
Lead compensation is adjustable via the parameter Lead compensation resistance
The total resistance (lead) must not exceed 320
Ω
with Pt100 and 3200
Ω
with Pt1000, KTY2X-6 or
resistance.
Thermo-
couple
Fe-Con J EN 60 584
-200 to +999 °C
0.4%/
≤
100ppm/°C
2
-
recognized
Fe-Con L DIN 43710
-200 to +900 °C
0.4%/
≤
100ppm/°C
2
-
recognized
NiCr-Ni K EN 60 584
-200 to +999 °C
0.4%/
≤
100ppm/°C
2
-
recognized
-10 to 60 mV
customer table
3
0.1%/
≤
100ppm/°C
3
-
recognized
For the voltage input (-10 to 60 mV), terminal temperature compensation for thermocouples can be used.
Internal terminal temperature compensation can be switched off through the setup program (0°C).