4 Technical data
4
Technical data
4.1
Input
4.1.1
Measuring range and accuracy
4.1.2
Reference conditions
Nominal width
Default settings for analog output
a,
b
Flow
max
a
DN 15 (Low-flow calibration)
0 to 30 l/min
60 l/min
DN 15
0 to 62.5 l/min
80 l/min
DN 20
0 to 105 l/min
210 l/min
DN 25
0 to 160 l/min
320 l/min
DN 32
0 to 260 l/min
520 l/min
a
Low flow cut off: 0.05 % of flow
max
.
b
Optionally expandable up to flow
max
.
X = Flow
max
(%)
Y = Measuring deviation from the
measured value ± (%)
Measuring deviation
a
Pulse output
≤ ±2.0 % of measured value ±0.1 % of flow
max
Current output
Like pulse output, additionally ≤ ±0.1 % of 20 mA
Voltage output
Like pulse output, additionally ≤ ±0.1 % of 10 V
Reproducibility
≤ ±0.5 % of measured value
a
±0.03 % of flow
max
Temperature drift
≤ ±0.15% of measured value
a
per 10 K temperature change (at -20 to +80° C)
Response time t
90
≤ 2 s
a
Under reference conditions.
1
2
3
4
5
6
7
8
9
10
0
0
10
20
40
60
80
100
30
50
70
90
%
%
X
Y
Measurement medium
Water
Medium temperature
23 °C (73 °F) ±5 K
Ambient temperature
23 °C (73 °F) ±5 K
Medium pressure
1 to 4 bar
Meter run
Horizontal installation, compliance with the required inlet and outlet sections