ME
12.3. List of commands
out commands:
Setting temperature values or parameters.
Command
Parameter
Response of circulator
out_mode_01
0
Use working temperature > SETPNT 1<
out_mode_01
1
Use working temperature > SETPNT 2<
out_mode_01
2
Use working temperature > SETPNT 3<
out_mode_04
0
Temperature control of internal bath.
out_mode_04
1
External control with Pt100 sensor.
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_mode_08
0
Set the control dynamics - aperiodic
out_mode_08
1
Set the control dynamics - standard
out_sp_00
xxx.xx
Set working temperature. „SETPNT 1“
out_sp_01
xxx.xx
Set working temperature. „SETPNT 2“
out_sp_02
xxx.xx
Set working temperature. „SETPNT 3“
out_sp_03
xxx.xx
Set high temperature warning limit „OVERTMP“
out_sp_04
xxx.xx
Set low temperature warning limit „SUBTMP“
out_sp_07
x
Set the pump pressure stage. (1 ... 4)
out_par_04
x.x
CoSpeed 0 ... 5.0 Band limit during external control.
Setting the maximum difference between the temperatures
in the internal bath and external system.
out_par_06
xxx
Xp control parameter of the internal controller.
0.1 ... 99.9
out_par_07
xxx
Tn control parameter of the internal controller.
3 ... 9999
out_par_08
xxx
Tv control parameter of the internal controller.
0 ... 999
out_par_09
xxx
Xp control parameter of the cascade controller. 0.1 ... 99.9
out_par_10
xxx
Proportional portion of the cascade controller.
1 ... 99.9
out_par_11
xxx
Tn control parameter of the cascade controller. 3 ... 9999
out_par_12
xxx
Tv control parameter of the cascade controller. 0 ... 999
out_par_13
xxx
Maximum internal temperature of the cascade controller.
out_par_14
xxx
Minimum internal temperature of the cascade controller.
out_par_15
xxx
Band limit (upper) 0 ... 200 °C
out_par_16
xxx
Band limit (lower)
0 ... 200 °C
61