![Julabo EW-12151-31 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/julabo/ew-12151-31/ew-12151-31_operating-manual_2030702010.webp)
Operator responsibility – Safety recommendations
F33-ME
F34-ME
Working temperature range
°C
-30 ... 200
-30 ... 150
Temperature stability
°C
±0,01
±0,01
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20 -30
0.5 0.32 0.12 0.03
+20 0 -20 -30
0.45 0.32 0.14 0.3
Refrigerant
R134a
R134a
Overall dimensions
(WxDxH)
cm
36x46x69
38x58x62
Bath opening
(WxL)
cm
23x14
24x30
Bath depth
cm
20
15
Filling volume
from ... to
liters
12 ... 16
14 ... 20
Weight
kg
44
42
Mains power connection
230 V/50 Hz
V/ Hz 230 / 50
230 / 50
Current draw
(at 230 V)
A
12
12
Current draw
GB, CH (at 230 V)
A
<9 / 2>
<9 / 2>
Mains power connection
V/ Hz 208…230 / 60
208…230 / 60
Current draw
(at 208 V / 230 V)
A
12
13
Mains power connection
V/ Hz 115 / 60
115 / 60
Current draw
(at 115 V)
A
15
14
Mains power connection
V/ Hz 100 / 60
----
Current draw
(at 100 V)
A
15
----
FP40-ME
FP50-ME
FPW50-ME
Working temperature range
°C
-40 ... 200
-50 ... 200
Temperature stability
°C
±0,01
±0,01
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20 -30
0.68 0.5 0.32 0.04
+20 0 -20 -30 -40
0.9 0.8 0.5 0.32 0.16
Refrigerant
R404A
R404A or R507
Overall dimensions
(WxDxH)
cm
37x46x69
42x49x70
Bath opening
(WxL)
cm
23x14
18x12
Bath depth
cm
20
15
Filling volume
from ... to
liters
9 ... 16
5.5 … 8
Weight
kg
48
55
Mains power connection
V/ Hz 230 / 50
230 / 50
Current draw
(at 230 V)
A
13
14
Current draw
GB, CH (at 230 V)
A
<9 / 3>
<9 / 4>
Mains power connection
V/ Hz 230 / 60
230 / 60
Current draw
(at 230 V)
A
13
14
All measurements have been carried out at:
rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.