![Julabo EW-12151-31 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/julabo/ew-12151-31/ew-12151-31_operating-manual_2030702021.webp)
ME
•
Chloric water can cause pitting corrosion.
•
Distilled and deionized water is unsuitable. Their special properties cause
corrosion in the bath, even in stainless steel.
Recommended bath fluids:
Bath fluid
Temperature range
soft/decalcified water
5 °C to 80 °C
mixture water/glycol, mixture 1:1
-20°C to 50°C
See website for list of recommended bath fluids.
ATTENTION: The maximum permissible viscosity is 50 mm
2
/s.
Caution:
Fire or other dangers when using bath fluids that are not recommended:
Please contact JULABO before using other than recommended bath fluids.
Use only nonacidic and noncorrosive bath fluids.
JULABO assumes no liability for damage caused by the selection of an unsuitable
bath liquid.
Unsuitable bath fluids are fluids which, e.g.,
•
are highly viscous
(much higher than recommended at the respective working temperature)
•
have a low viscosity and have creep characteristics
•
have corrosive characteristics or
•
tend to crack.
•
No liability for use of other bath fluids!
5.3.
Temperature application to external systems
Caution:
Securely attach all tubing to prevent slipping.
If the circulator is operated without external system, close the pump connector (24a)
with the cap nut.
feed
return
The circulator is used for temperature application to external,
closed systems (loop circuit) with simultaneous temperature
application in the circulator bath.
Connecting the external system
•
Unscrew the collar nuts from the pump connector (24a).
•
Slide the tubing onto the pump connector for feed (24a) and
return flow (24b) and secure with hose clamps.
21