Operator responsibility – Safety recommendations 

 

2.4. Technical 

specifications 

 

 

ED 

 

Temperature selection 

 

digital 

 

Temperature indication 

 

LED 

 

Resolution 

°C 

0.1 

 

Temperature control 

 

PID1 

 

Heater wattage  (at 230 V) 

kW 

2,0 

 

Heater wattage  (at 115V) 

kW 

1,0 

 

Circulating pump: 

 

 

 

 

discharge, max.at 0 bar 

l/min 

15 

 

 

pressure, max. at 0 l 

bar 

0.35 

 

Ambient temperature 

°C 

5 ... 40 

 

 

 

 

 

Mains power connection 230 V/50 Hz 

V/ Hz 

190-253 / 50 

 

Current input (at 230 V) 

 

 

Mains power connection 230 V/60 Hz 

V/ Hz 

190-253 / 60 

 

Current input (at 208 V / 230 V) 

8 / 9 

 

Mains power connection 115 V/60 Hz 

V/ Hz 

103-127 / 60 

 

Current input (at 115 V) 

 

 

Mains power connection 100 V/50-60 Hz 

V/ Hz 

90-110 / 50-60 

 

Current input (at 100 V) 

 

 

 

 

 

 

All measurements have been carried out at: 

 

rated voltage and frequency 

operating temperature: 70 °C  

ambient temperature: 20 °C  

bath fluid: water 

 

Technical changes without prior notification reserved. 
 
 

0pen Bath Circulators 
Bath tanks 

 

ED-5A 

Plexiglas 

®

 

ED-5M  

Makrolon 

®

 

ED-5 
Stainless 
steel 

Working temperature range 

 

20 ... 60 

20 ... 100 

20 ... 100 

Temperature stability 

 

±0,03 

±0,03 

±0,03 

 

 

 

 

 

Bath opening (WxL) 

cm 

12x24 

12x24 

15x15 

Bath 

depth 

cm 

15 15 15 

Filling volume 

liters 

3,5 ... 5 

3,5 ... 5 

3 ... 4,5 

Overall dimensions (WxDxH) 

cm 14x40x35 14x40x35 17x33x36 

Weight 

kg 

5,3 

5,2 

7,0 

 

Содержание ED-5

Страница 1: ...270 doc 02 02 11 English Operating manual Open Bath Circulators ED 5A ED 5M ED 5 JULABO Labortechnik GmbH 77960 Seelbach Germany 49 0 7823 51 0 49 0 7823 24 91 info julabo de www julabo de 1 951 0270...

Страница 2: ...ndustry are developed produced and distributed according to the requirements of ISO 9001 2008 Certificate Registration No 01 100044846 Unpacking and inspecting Unpack the circulator and accessories an...

Страница 3: ...ols and functional elements 13 5 Preparations 14 5 1 Installation 14 5 2 Bath fluids 14 5 3 Filling Draining 15 5 4 Temperature application to external systems 16 5 5 Tubing 16 5 6 Adjusting the pump...

Страница 4: ...temperature display The PID temperature control adapts the heat supplied to the thermal requirements of the bath Safety installations conforming to IEC 61010 2 010 The excess temperature protection is...

Страница 5: ...ive affects on the instrument being operated in the same mains This unit is designed for operation in a controlled electromagnetic environment This means that transmitting devices e g cellular phones...

Страница 6: ...g label W26 Colors yellow black Hot surface warning The label is put on by JULABO 2 1 Disposal Do not dispose of the unit with household waste However over the long operating period of the unit dispos...

Страница 7: ...accordance with appropriate operating instructions for a period of ONE YEAR Extension of the warranty period free of charge With the 1PLUS warranty the user receives a free of charge extension to the...

Страница 8: ...A 8 9 Mains power connection 115 V 60 Hz V Hz 103 127 60 Current input at 115 V A 9 Mains power connection 100 V 50 60 Hz V Hz 90 110 50 60 Current input at 100 V A 9 All measurements have been carrie...

Страница 9: ...for storage and transportation Air humidity Max rel humidity 80 for temperatures up to 31 C linear decrease down to 50 relative humidity at a temperature of 40 C Max mains fluctuation of 10 are permi...

Страница 10: ...oided slight or minor injuries could result A warning of possible property damage may also be contained in the text Notice Describes a possibly harmful situation If this is not avoided the product or...

Страница 11: ...id prior to draining by switching the unit on for a short moment for example Observe the limited working temperature range when using plastic bath tanks Employ suitable connecting tubing Make sure tha...

Страница 12: ...ed and installed Note sharp edges hot surfaces in operation moving machine parts etc Do dangerous steams or gases arise when heating Is an exhaust needed when working What to do when a dangerous subst...

Страница 13: ...shortly for step by step changes Keep the keys pressed for fast change of setpoint 3 OK key store 4 LED temperature display menu indication 5 Control indicator Heating 7 Control indicator Alarm 7 Adju...

Страница 14: ...bath tank with the bath cover B 5 2 Bath fluids Caution No liability for use of other bath fluids Do not use flammable bath fluids Water The quality of water depends on local conditions Due to the hi...

Страница 15: ...om time to time Pump and heater must always be fully covered with the bath fluid The instrument is not suited for unsupervised continuous operation Filling Take care that no liquid enters the interior...

Страница 16: ...onto the pump connectors for feed 24a and return flow 24b Secure the tubing with tube clamps ED 5A ED 5M ED 5 5 5 Tubing Recommended tubing Order No Length Temperature range 8930008 1 m CR tubing 8 m...

Страница 17: ...eck the tubing for material defects e g for cracks Preventive maintenance Replace the tubing from time to time 5 6 Adjusting the pump flow 1 2 The pump flow is pre adjusted in the factory and can be m...

Страница 18: ...upply and lead the tap water in a sink through the return connector 25b Cooling coil ED 5A ED 5M Cooling coil ED 5 LA specific water flow rate of 45 ml minute is sufficient to compensate for the chara...

Страница 19: ...the identification plate 6 2 Switching on Start Stop Caution Observe the limited working temperature range when using plastic bath tanks Bath tanks made of Plexiglas up to 60 C Bath tanks made of Makr...

Страница 20: ...ated by OFF on the LED temperature display A complete shutdown of the main functional elements such as heater and circulating pump is effected simultaneously Should such a safety standard not be requi...

Страница 21: ...rature of the bath fluid has reached the safety temperature a complete shutdown of the heater and pump is effected The alarm is indicated by optical and audible signals continuous tone and on the LED...

Страница 22: ...Cable of the working temperature sensor interrupted or short circuited Defect of the working or excess temperature sensor Working temperature and excess temperature sensors report a temperature diffe...

Страница 23: ...he circulator is designed for continuous operation under normal conditions Periodic maintenance is not required The tank should be filled only with a bath fluid recommended by JULABO To avoid contamin...

Отзывы: