ED
5
Persons:
The operator is responsible for the qualification of the personnel operating the units.
The operator should be constantly informed about the dangers involved with their job activities
as well as preventive actions.
Make sure all persons expected to carry out operation, installation and maintenance of the unit
read and understand the safety information and operating instructions.
If you have any questions concerning the operation of your unit or the information in this
manual, please contact us!
Contact:
JULABO Labortechnik GmbH
Eisenbahnstraße 45
77960 Seelbach / Germany
Tel. +49 (0) 7823 / 51-0
Fax +49 (0) 7823 / 24 91 www.julabo.de
Handling:
Nevertheless, avoid strikes to the housing, vibrations, damages to the keypad foil (keys,
display) or contamination.
Make sure the product is regularly checked for proper condition.
Regularly check (at least every 2 years) the proper condition of the mandatory, warning,
prohibition and safety labels.
Take care that the mains supply features a low impedance to avoid any negative affects on
the instrument being operated in the same mains.
This unit is designed for operation in a controlled electromagnetic environment. This
means that transmitting devices (e.g. cellular phones) should not be used in the immediate
vicinity.
Magnetic radiation may influence other units with components susceptible to magnetic
fields (e.g. a monitor). We recommend to keep a minimum distance of 1 m.
Permissible ambient temperature: max. 40 °C, min. 5 °C.
Permissible relative air humidity: 50 % (40 °C).
Do not store in an aggressive atmosphere. Protect from contaminations.
Do not expose to sunlight.
Appropriate operation:
Only qualified personnel is authorized to perform configuration, installation, maintenance and
repairs of the circulator.
Routine operation can also be carried out by untrained personnel who should however be
instructed by trained personnel.
Use:
The bath may not be filled with flammable materials. Fire hazard!
Only use recommended materials (bath fluids). Only use recommended materials (bath fluids).
Only use non-acid and non corroding materials.
When using hazardous materials or materials that could become hazardous, the operator
must affix the enclosed safety labels (1 + 2) to the front of the unit so they are highly visible: