![Julabo CORIO CP-1000F Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/julabo/corio-cp-1000f/corio-cp-1000f_original-operating-manual_2030700055.webp)
Operation
55
8.10 Adjusting the temperature sensor (ATC)
For physical reasons, there can be a temperature difference in the bath tank
between the temperature sensor and a defined, more remote point within the
bath fluid volume. As a result, the measured temperature deviates slightly from
the actual bath temperature. Adjustment of the temperature sensor can increase
accuracy of the temperature control application.
The adjustment of a temperature sensor requires the determination of the
temperature difference between the temperature sensor and the reference
thermometer. Three adjustment options are available:
• 1-point adjustment: When a sample is tempered to a certain setpoint. The
adjustment curve is displaced parallel to the original curve by the reference
amount.
• 2-point adjustment: When the temperature is controlled between two
setpoints. The slope of the temperature curve is adjusted between two
points.
• 3-point adjustment: When temperature is controlled within a temperature
range. This results in an arched temperature curve.
8.10.1 Adjust internal temperature sensor
This section describes how to adjust the internal temperature sensor of the
device.
The bath tank is filled.
The device is switched on.
The ATC function is disabled.
1. Hang the calibrated thermometer in the bath tank and place the bath lid on
top.
2. Press the
[MENU]
key.
3. Use the arrow keys to select the menu option
<Atc>
and confirm with
[OK]
.
4. Use the arrow keys to select the menu option
<tYPE>
and confirm with
[OK]
.
5. Use the arrow keys to select the number of correction points and confirm
with
[OK]
.
6. Exit the menu and start the temperature control application with the desired
setpoint temperature.
When the setpoint is reached, allow the temperature to stabilize for several
minutes.
The more stable the temperature in the bath tank, the more precise the
adjustment result.