Safety
14
4.3 Safety function
Technical protective devices provide for safe operation. If a safety function is
triggered, the operator is alerted with a message on the display and an acoustic
signal.
Adjustable high temperature cut-off
The high temperature cut-off prevents overheating of the heater.
• If the measured temperature rises above the set protective temperature, an
error message is shown on the display. The pump and heater are switched
off. A restart is required.
Overheating protection
The overheating protection prevents overheating of the heater.
• The protective mechanism is triggered when the device recognizes a
temperature difference of more than 20 K between the working
temperature sensor and the safety temperature sensor. Am error message
appears on the display. A restart is required.
Low liquid level protection
A level switch recognizes when the bath fluid fill level in the bath tank is too
low. The device has a two-stage warming system to prevent overheating of the
heater or dry running of the pump.
• The low liquid level warning is triggered when the liquid level in the bath
tank is about half the full level. A warning message appears on the display.
The bath fluid should be topped up.
• The low liquid level alarm is triggered when the float reaches its lower limit
stop. The device switches off the pump and heater. A continuous signal
tone sounds. A warning message appears on the display. A restart is
required.