Cryo-Compact Circulators
75
Command
Parameter Response of Cryo-Compact Circulator
IN_PAR_11
none
Tn control parameter of the cascade controller.
IN_PAR_12
none
Tv control parameter of the cascade controller.
IN_PAR_13
none
Adjusted maximum internal temperature of the cascade controller.
IN_PAR_14
none
Adjusted minimum internal temperature of the cascade controller.
IN_PAR_15
none
Band limit (upper)
IN_PAR_16
none
Band limit (lower)
IN_MODE_01
none
Selected setpoint:
0 = Setpoint 1
1 = Setpoint 2
2 = Setpoint 3
IN_MODE_02
none
Selftuning type:
0 = Selftuning „off“
1 = Selftuning „once“
2 = Selftuning „alway“
IN_MODE_03
none
Type of the external programmer input:
0 = Voltage 0 V to 10 V
1 = Current 0 mA to 20 mA
IN_MODE_04
none
Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
IN_MODE_05
none
Cryo-Compact Circulator in Stop/Start condition:
0 = Stop
1 = Start
IN_MODE_08
none
Adjusted control dynamics
0 = aperiodic
1 = standard
IN_HIL_00
none
Max. cooling power (%).
IN_HIL_01
none
Max. heating power (%).
11.4. Status messages
Status messages
Description
00 MANUAL STOP
Cryo-Compact Circulator in „OFF“ state.
01 MANUAL START
Cryo-Compact Circulator in keypad control mode.
02 REMOTE STOP
Cryo-Compact Circulator in „r OFF“ state.
03 REMOTE START
Cryo-Compact Circulator in remote control mode.