Menu functions
44
Examples:
-OFF-
S 25.00
E -----
Setpoint3: 80.00°C
IntAct : 80.00°C
Power : 2%
Control : intern
Setpoint
Setpoint1: 25.00°C
>Setpoint2: 58.00°C
Setpoint3: 85.00°C
Setpoint2: 58.00°C
IntAct : 60.00°C
Power : 0%
Control : intern
60.00
S 58.00
E -----
Profile End Stdby / SetpPG / Setp 1 / Setp 2 / Setp 3
This parameter is used to determine if the unit switches to the -OFF-
condition at the end of a profile, or if temperature control is continued
and the working temperature setpoint that is be used in such case.
Stdby
– If the end of the profile is reached, the Cryo-Compact
Circulator switches to the -OFF- condition. The VFD display
indicates the working temperature setpoint, that was displayed in line
2 already before the program was started (example: S 25.00 °C ).
SetpPG
– (Setpoint of the
P
ro
G
rammer) At the end of the profile, the
setpoint of the last program section is written to the position of
setpoint 3 (example: 80.00 °C) and indicated in line 1 of the LCD
display. The Cryo-Compact Circulator operates and controls this
temperature until the Start/Stop key
is pressed or a new
profile is started.
Setp 1, Setp 2, Setp 3
– („
Setp
“ short for
Setpoint
)
Before starting the programmer, a temperature value is set for one of
the three setpoints in the
menu.
(example: Setpoint 2 58.00 °C)
The Cryo-Compact Circulator operates and controls this
temperature at the end of the profile until the Start/Stop key
is
pressed or a new profile is started.
Start
no / yes
(manual start)
or
time
(via integrated timer)
When selecting the parameter
time
, a new submenu is called up for
entry of the start time.
A flashing segment indicates that a start time needs to be entered.
hour.min
Start time
Day.Mon
day and month
Year
year
Set each entry with enter
.
Start
no / yes
A flashing line indicates that the parameter „yes“ needs to be entered.
Press the key to select the parameter and press enter
.
Starttime: 06:00
Date : 15:12
*** wait ***
Acttime 02:34:45
The temperature system switches to waiting mode and a flashing line
„wait“ appears on the LCD DIALOG-DISPLAY. The start time and actual
time are permanently indicated on the display.