9 Alarm indicators
Overtemperature indicator (2). The safety thermostat has disconnected the heater:
Possible cause 1:
- The equipment has been switched on without or with not enough water.
Possible solution:
1. Add water untill the heater element cover rack, as indicated in chapter 8.1.
2. Press the safety thermostat restart button (19) and begin the sterilization process
once again.
Possible cause 2:
- The valve is not closed (12) and after the purge phase, the water in the sterilization
chamber has been evaporated.
Possible solution:
1. Do the same as in cause 1 and be sure, when initiating the sterilization process,
to set the valve (12) to the «Closed» position after purging. See chapter 8.3.
Open lid indicator (3)
Possible cause:
- The indicator is on while the lid is open and must be off when closing it. If it does
not switch off or on when operating, carefully depressurize the autoclave and
contact an authorized technical service of JP Selecta, SA.
2
3
10
5
8
6
7
4
9
1
10 Spare Parts
07001
Network connection cable
13007
Restart and heater contactors
15480
Fuse carrier
15595
Fuse
16078
Pressure gauge
16212
Pressure switch 2 bar (134ºC)
16220
Pressure switch 1 bar (121ºC)
20061
Temperature selector
20096
Main switch
22003
Amber indicator light
22004
Red indicator light
22063
Green indicator light
34057
Three channels valve
34202
Safety valve
37025
Timer
- Liquids sterilization: do not open the valve when it cools down till the pressure
gauge (1) indicates 0 bar. Then, set the valve (12) to the «Steam» position.
In both cases, please open the autoclave lid when the pressure gauge indicates 0
bar, as explained in chapter 8.2.