Additional Notes

11.Do not place the blood pressure monitor and cuff at will. It will cause asphyxiation if the child 
swallows or twine around his neck.
12.The cuff and the case of the blood pressure monitor have been tested for biocompatibility and 
do not contain allergenic or harmful materials.Please stop using it if allergy occurs during use.
13.Warning:
Non-professionals do not modify the equipment, otherwise it will make the equipment 
measurement is not accurate.
14.Warning:
Do not expose the equipment for a long time, otherwise it will reduce the performance of the 
equipment.
15.Warning:
This device is not used for children and pets
16.Clean:
The equipment can be cleaned by lay operator according to the cleaning procedures in the 
instructions
17.Warning:
Do not use a damaged cuff for blood pressure measurement.
18.Warning:
When measuring with the cuff, if the tester feels seriously uncomfortable, press the button of the 
blood pressure monitor to deflate the cuff, or remove the cuff directly from the arm.
19.Warning:
If an unexpected reading occurs, the operator can take several more measurements and consult a 
doctor.
20.Warning:
This equipment is used outside the specified environment, may damage the equipment, and may be 
inaccurate measurement.

Correct Disposal of This Product

(Waste Electrical & Electronic Equipment) 

This marking shown on the product indicates that it should not be disposed with other 
household waste at the end of its life. To prevent potential harm to the environment or to 
human health, please separate this product from other types of wastes and recycle it 
responsibly.  When disposing this type of product, contact the retailer where product was 
purchased or contact your local government office for details regarding how this item can 
be disposed in an environmentally safe recycling center.
Business users should contact their supplier and check the terms and conditions of the 
purchasing agreement. This product should not be mixed with other commercial wastes 
for disposal. This product is free of hazardous materials.

37

Important Instructions Before Use
1. WARNING: Use of this equipment adjacent to or stacked with other equipment should be 
avoided because it could result in improper operation. If such use is necessary, this equipment and 
the other equipment should be observed to verify that they are operating normally.
2.WARNING: PORTABLE RF communications equipment (including peripherals such as antenna 
cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of  Arm-
type Fully Automatic Digital Blood Pressure Monitor, including cables specified by the 
MANUFACTURER. Otherwise, degradation of the performance of this equipment could result.
3.The software identifier refer to the software evaluation report , and the file code is

JYRJ201006001.

4.

verify manometer pressure accuracy:

In the power down state, press and hold the " START/STOP" button,  and theninstall 
the batteries.  Until the LCD screen is full, release the " START/STOP" button. 
When the LCD screen displays the  double zero, the bloodpressure meter is in static state.
At this point, 500ml gas capacity, calibrated standard pressure gauge and manual pressure device 
can be connected to the sphygmomanometer through the sleeve interface of the 
sphygmomanometer, and manual pressure can be applied to the effective display range of the 
sphygmomanometer, and then the difference between the reading of the sphygmomanometer and 
that of the standard pressure gauge can be compared.This mode can be used to verify manometer 
pressure accuracy.
5.Contraindications:
Product is not intended for infants or individuals who cannot express their intentions.
6.Intended Use
The digital blood pressure monitor are reusable for clinical and home use and are non-invasive 
blood pressure measurement systems designed to measure the systolic and diastolic blood pressure 
and pulse rate of adolescents and adults individual by using a non-invasive technique, which is a 
well-known technique in the market called the “oscillometric method”.
it can measure the systolic blood pressure, diastolic blood pressure and pulse rated on 
up-arm,and the device is reusable for clinical or home use.
7.The patient is the operator:
the PATIENT is an intended OPERATOR.
the PATIENT Do not carry out other maintenance operations except to replace the battery.
8.WARNING:
 Do not modify this equipment without authorization of the manufacturer.
9. ESSENTIAL PERFORMANCE Maintenance advice:
Pressure calibration will be carried out when this product leaves the factory. Patients can use the 
method described in the section "Verify Manometer Pressure Accuracy" to verify the accuracy. 
If the accuracy deviation is large, please contact the manufacturer to recalibration.
10.Mechanical strength and resistance to heatThe resistance to heat will be retained by device 
during the EXPECTED SERVICE LIFE of the ME EQUIPMENT.

Additional Notes

Recommended separation distances between portable and mobile RF 
communications equipment and the device

The device is intended for use in an electromagnetic environment in which 
radiated therefore disturbances are controlled. The customer or the user of the 
device can help prevent electromagnetic interference by maintaining a 
minimum distance between portable and mobile RF communications 
equipment (transmitters) and the device as recommended below, according to
the maximum output power of the communications equipment.

Rated maximum
output power of
transmitter
           
           W

Table 4

Separation distance according to frequency of transmitter
                                                   m

80 MHz to 800 MHz

800 MHz to 2.7 GHz
 

0.01

0.1

1

10

100

0.12

0.38

1.2

3.8

12

0.23

0.73

2.3

7.3

23

For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.

NOTE1 At 80 MHz and 800 MHz, the separation distance for th
frequency range applies.

NOTE2 These guidelines may not apply in all situations. Electromagnetic 
propagation is affected by absorption and reflection from structures, objects 
and people.

e higher

 

Electromagnetic Compatibility Information

36

Table 3

Electromagnetic Compatibility Information

35

Guidance and declaration of manufacturer-electromagnetic immunity

Nowadays, many RF wireless equipments have being used in various healthcare
locations  where  medical  equipment  and/or  systems  are  used.  When  they  are 
used  in  close  proximity  to  medical  equipment  and/or  systems,  the  medical 
equipment  and/or  systems’  basic  safety  and  essential  performance  may  be 
affected. Arm-type  Fully Automatic  Digital  Blood  Pressure  Monitor has  been 
tested  with  the  immunity  test  level  in  the  below  table  and  meet  the  related 
requirements  of  IEC  60601-1-2:2014.  The  customer  and/or  user  should  help 
keep a minimum distance between RF wireless communications equipment and 
this medical equipment and/or systems as recommended below.

 

     Service

1970

     Immunity
     test level
         (V/m)

    Distance
        (m)

 Band
(MHz)

0.3

     Test 
frequency
    (MHz)

     Modulation

     Maximum 
        power
           (W)

     TETRA 400

    0.2

     9

5785

5500

5240

2400-2570

     2450

     800-960

    
   28

0.3

     2

704-787

0.2

710

0.3

     27

     0.3

    1.8

     450

9

28

0.3

   2

745

780

810

870

     

GMRS 460

        FRS 460

     

FM

±

 5 kHz deviation

1 kHz sine

2

28

 430-470

        Pulse 
     modulation
         18Hz

   380-390

     385

930

1720

1700-1990

2

1845

0.3

    
   28

0.3

   

GSM 1800;

  CDMA 1900;
  GSM 1900;
    DECT;
 LTE Band 1, 3,
   4, 25; UMTS

    Pulse
 modulation
      18Hz

GSM 800/900,
  TETRA 800,
   iDEN 820,
  CDMA 850,
  LTE Band 5

     Pulse
 modulation
     217Hz

      LTE 
      Band 
      13, 17

Pulse modulation
       217Hz

   WLAN 
     802.11
       a/n

        Pulse 
    modulation
        217Hz

 

Bluetooth,WLAN,

802.11 b/g/n,RFID
 2450,LTE Band 7

    Pulse 
modulation
     217Hz

    5100-5800

Содержание DBP-6281B

Страница 1: ...peration 1 Do not drop the unit Protect it from sudden jars or shocks 2 Do not insert foreign objects into any openings 3 Do not attempt to disassemble the unit 4 Do not crush the pressure cuff 5 If the unit has been stored at temperatures below 0 oC leave it in a warm place for about 15 minutes before using it Otherwise the cuff may not inflate properly 6 If the unit has been stored at temperatur...

Страница 2: ...ce or helicopter 30 Contains small parts that may cause a chocking hazard if swallowed by infants 31 Please align the polarities of each battery with the ve and ve signs imprinted on the battery housing when you replace the batteries 32 Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this equ...

Страница 3: ...k Start 1 Install batteries 2 Insert cuff air plug into the behind side of monitor unit See Figure B See Figure A 3 Remove thick clothing from the arm area 4 Rest for several minutes prior to testing Sit down in a quietplace comfortably back and arm support on a desk or table withyour legs uncrossed your arm resting on a firm and your feet flat on the floor See Figure C 5 Apply cuff to your left a...

Страница 4: ... enter volume setting mode Set the voice volume by adjusting the MEM button There are six volume levels While in any setting mode press START STOP button to turn the unit off All information will be saved Note If unit is left on and not in use for 3 minutes it will automatically save all information and shut off Unit Operation System Settings With power off press GROUP1 button or to activate Syste...

Страница 5: ...ldest test result in memory can be viewed by pressing and holding the MEM button Upon activating test results you can press the MEM button to scroll through all test results stored in memory To check the average results from other groups please press corresponding GROUP1 button or GROUP2 button With power off press the MEM button to activate screen display After the unit performs a self diagnosis ...

Страница 6: ...2483 5 MHz Modulation GFSK Antenna gain 0 5dBi Troubleshooting Problem Solution displayed Connection failure Data is not being transmitted Cuff is too tight or not properly positioned on the arm Firmly reposition cuff approximately above the elbow joint See Page 12 1 2cm 1 2 Inaccurate test results due to body movement or monitor movement Sit in a relaxed position with arm placed near heart Avoid ...

Страница 7: ...essure varies throughout the course of a day Many factors including diet stress cuff placement etc may affect an individual s blood pressure Q Should I apply the cuff to the left or right arm What is the difference A Either arm can be used when testing however when comparing results the same arm should be used Testing on your left arm may provide more accurate results as it is located closer to yo...

Страница 8: ...rating Environment Automatic Pressurization LCD Digital Display Size 87 5mm X 85mm 3 44 x 3 34 Humidity Temperature 10 40 50 104 15 93 RH Storage Environment Humidity Classification Ingress Protection rating Battery Shelf life Battery Storage Temperature Temperature Internal Powered Equipment Type BF Cuff is the Applied Part IP 20 Indoor Used Only 60 months 25 55 13 131 25 55 13 131 Measurement Ra...

Страница 9: ...al environment Group 1 class B Table 2 continued Guidance and declaration of manufacturer electromagnetic immunity The device is intended for use in the electromagnetic environment specified below The customer or the user of the device should assure that it is used in such an environm ent Electromagnetic environment guidance Compliance level IEC 60601 test level IMMUNITY test Electromagnetic Compa...

Страница 10: ...e use and are non invasive blood pressure measurement systems designed to measure the systolic and diastolic blood pressure and pulse rate of adolescents and adults individual by using a non invasive technique which is a well known technique in the market called the oscillometric method it can measure the systolic blood pressure diastolic blood pressure and pulse rated on up arm and the device is ...

Отзывы: