24
JC-114
ULTRASONIC NEBULIZER
GB
EMISSION TESTS
Electrostatic
discharge
(ESD)
IEC/EN61000-4-2
Fast
transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN61000-4-5
Voltage dips, short
interruption and
power fluctuations
IEC/EN61000-4-11
Magnetic field
IEC/EN61000-4-8
Conducted
immunities
IEC/EN61000-4-6
Radiated
immunities
IEC/EN61000-4-3
COMPLIANCE
± 6kV in contact
± 8kV in air
± 2kV power supply
± 1kV differential
mode
< 5% UT for 0.5 cycles
40% UT for 05 cycles
70% UT for 25 cycles
< 5% UT for 5
seconds
3 A/m
3Vrms 150kHz
to 80MHz
(for devices that are
not life-supporting)
3 V/m 80 MHz a
2.5 GHz
(for devices that are
not life-equipment )
ELECTROMAGNETIC ENVIRONMENT GUIDE
Floor should be concrete, wood or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment
Mains power quality should be that of a
typical commercial or hospital environment
Power supply should be the same as the one
supplied in the commercial and hospital
environment. Whether the user needs a
continuous use, it will be necessary to
connect the unit to an UPS (uninterruptible
power supply)
Power frequency magnetic fields should be at
level characteristics of a typical location in a
typical commercial or hospital environment.
MANUFACTURER GUIDANCE AND DECLARATION – ELECTROMAGNETIC IMMUNITY
Nebulizer JC-114 can be used in the following electromagnetic environment. The customer
or the user of NEBULIZER JC-114 must make sure to use it in this environment.
Note: UT is the a.c. mains voltage prior to application of the test level
Specifications and designs are based on the latest information available at the time of printing
and subject to change without notice.