Jorvet J1509 Скачать руководство пользователя страница 8

8

EXAMPLE CALCULATION

OXYGEN SOURCE

A 10 kg patient at 12 breaths per minute on a desired 35% oxygen 
saturation: placed in a Smart Oxygen unit using an initial flow rate 
of 5 lpm.

u

  Use  Table  1  to  determine  the  approximate  amount  of  time  (in 

minutes) to reach the desired oxygen concentration based on 
Venturi  selection  and  flow  rate.  This  is  with  no  patient  inside.  
With the patient inside, it will take slightly longer to achieve the 
desired concentration.  

 

u

 

Example

 2.8 mins to achieve desired oxygen concentration

 Then, once concentration is achieved: 

u

  Use Table 2 to determine how much your patient’s tidal volume. 

u

 

Example

 0.10 lpm base on 10 ml/kg/min tidal volume

u

  Use Table 3 to determine the minimum oxygen flow rate needed 

to maintain the oxygen concentration base on the patient’s 
current respiratory rate.

 

u

 

Example

 2.4 lpm minimum oxygen flow rate (1/min) to maintain 

  oxygen

u

  Use the numbers from Tables 2 & 3 to determine the minimum 

flow rate neeed for the patient. This may change as the patient’s 
respiratory rate changes.

 

u

 

Example

 2.4 lpm + 0.10 lpm = 2.5 lpm minimum flow rate

    needed to maintain the desired oxygen concentration

Oxygen from an oxygen tank is the best source for the Smart 
Oxygen unit. Oxygen concentrators may be used but due to the 
lower PSI and oxygen concentration, the oxygen concentration 
inside the Smart Oxygen unit may not be what is expected or stated 
in the instructions. Using an oxygen analyzer inside the cage can 
help the user ensure the correct concentration is achieved when 
using an oxygen concentrator as the oxygen source.

Содержание J1509

Страница 1: ...SMART OXYGEN CAGE J1509 USER MANUAL Oxygen Therapy An important modality for critically ill patients...

Страница 2: ...gen therapy and nebulization Figure 1 SMART OXYGEN OVERVIEW CONTENTS A I J A B C D E G F u A u B u C u D u E u F u G u H u I u J 2x Elbow Outlets w Cups Double Male Connector Venturi Oxygen Blank Opaq...

Страница 3: ...outlet Connect one end of the tubing D directly to your oxygen supply u Connect the green Venturi E to the other end of the tubing and attach it to the Smart Oxygen inlet port via the double male conn...

Страница 4: ...enturi As the oxygen passes though the Venturi it draws in room air which then mixes with the oxygen The larger the size of holes in the Venturi the more room air is drawn in The resulting mixture the...

Страница 5: ...the oxygen at the required flow rate Ensure the outlet port is uncapped to prevent pressure and CO2 from building up in the system Attach waste gas anesthesia scavenging tubing if necessary u Allow t...

Страница 6: ...ttach the double male connector B to one port and the other to your waste gas anesthesia scavenging tubing u Connect the tubing to your oxygen source u If using an anesthesia machine Connect the blank...

Страница 7: ...sh the unit with oxygen before lifting the lid Failure to do so may result in medication or gas escaping into the procedure area Patient breathing rate should be monitored as it may change due to anxi...

Страница 8: ...determine the minimum oxygen flow rate needed to maintain the oxygen concentration base on the patient s current respiratory rate u Example 2 4 lpm minimum oxygen flow rate 1 min to maintain oxygen u...

Страница 9: ...f the display unit to select Celsius or Fahrenheit Humidity is displayed in percent The lowest and highest temperature humidity recorded over past 24 hours is also displayed along the right side of th...

Страница 10: ......

Страница 11: ......

Страница 12: ...Jorgensen Laboratories Inc Loveland CO 80538 800 525 5614 fax 970 663 5042 www jorvet com info jorvet com Professional Veterinary Equipment Supplier...

Отзывы: