To provide emergency oxygen therapy to a patient in respiratory 
distress:

Refer to Figure 1 on Page 2

u

  The Smart Oxygen unit has two ports 

A

 that may be used as inlet 

and outlet. Keep both ports uncapped and attach the double 
male connector 

B

 to one of the ports.

u

  Connect the tubing to your oxygen source.  

 

u

  If using an anesthesia machine - Connect the blank    

 

  Venturi (no holes) 

C

 to one end of the tubing 

D

 and attach    

  to the end of the anesthesia breathing circuit.

 

u

  If using a regulator/flowmeter with a male DISS 

    outlet – Connect the Christmas Tree Swivel Adaptor 

G

 to  

 

  one end of the tubing 

D

 and attach to your oxygen supply.

 

u

  If using a regulator/flowmeter with a barbed outlet –     

 

  Connect one end of the tubing 

D

 directly to your oxygen  

 

  supply.  

u

 Connect the green Venturi 

E

 to the other end of the tubing 

and attach it to the Smart Oxygen inlet port via the double 
male connector 

B

. The opposite port is considered the outlet 

now. 

Do not cap the outlet port while a patient is inside the 

Smart Oxygen cage!

 If necessary, attach waste gas anesthetic 

scavenging system tubing to the outlet port.  

u

  Initially, an oxygen flow rate of 4 L/min for a patient of up to 10 

kg is recommended.

u

  Place patient inside the Smart Oxygen unit, close all the zips, turn 

on the oxygen and maintain constant patient observation. Allow 
the  patient  to  rest  briefly  in  the  oxygen-enriched  environment 
before further investigation or handling. For more details in 
the correct use of the Smart Oxygen unit, you must read all 
instructions contained within this manual. 

QUICK START

Содержание J1509

Страница 1: ...SMART OXYGEN CAGE J1509 USER MANUAL Oxygen Therapy An important modality for critically ill patients...

Страница 2: ...gen therapy and nebulization Figure 1 SMART OXYGEN OVERVIEW CONTENTS A I J A B C D E G F u A u B u C u D u E u F u G u H u I u J 2x Elbow Outlets w Cups Double Male Connector Venturi Oxygen Blank Opaq...

Страница 3: ...outlet Connect one end of the tubing D directly to your oxygen supply u Connect the green Venturi E to the other end of the tubing and attach it to the Smart Oxygen inlet port via the double male conn...

Страница 4: ...enturi As the oxygen passes though the Venturi it draws in room air which then mixes with the oxygen The larger the size of holes in the Venturi the more room air is drawn in The resulting mixture the...

Страница 5: ...the oxygen at the required flow rate Ensure the outlet port is uncapped to prevent pressure and CO2 from building up in the system Attach waste gas anesthesia scavenging tubing if necessary u Allow t...

Страница 6: ...ttach the double male connector B to one port and the other to your waste gas anesthesia scavenging tubing u Connect the tubing to your oxygen source u If using an anesthesia machine Connect the blank...

Страница 7: ...sh the unit with oxygen before lifting the lid Failure to do so may result in medication or gas escaping into the procedure area Patient breathing rate should be monitored as it may change due to anxi...

Страница 8: ...determine the minimum oxygen flow rate needed to maintain the oxygen concentration base on the patient s current respiratory rate u Example 2 4 lpm minimum oxygen flow rate 1 min to maintain oxygen u...

Страница 9: ...f the display unit to select Celsius or Fahrenheit Humidity is displayed in percent The lowest and highest temperature humidity recorded over past 24 hours is also displayed along the right side of th...

Страница 10: ......

Страница 11: ......

Страница 12: ...Jorgensen Laboratories Inc Loveland CO 80538 800 525 5614 fax 970 663 5042 www jorvet com info jorvet com Professional Veterinary Equipment Supplier...

Отзывы: