Jorvet J1509 Скачать руководство пользователя страница 4

4

Supplementary oxygen should be provided immediately to 
patients with dyspnea or cyanosis. Improved blood oxygenation 
of the patient may also be achieved when using the unit pre and 
post anesthesia. The oxygen percentage within the Smart oxygen 
unit is determined by the extent of filling it with air and oxygen via 
the Venturi. Patients in respiratory crisis should be given a high 
percentage of oxygen initially and then the concentration gradually 
reduced to the lowest level at which the patient can breathe 
comfortably. Opening the zips and carrier lid will drop the oxygen 
level to that of room air almost immediately.

Venturi

This kit contains 5 Venturi in various colors. The colors represent 
an air to oxygen ratio. Oxygen from your supply flows through the 
tubing into the attached Venturi. As the oxygen passes though the 
Venturi, it draws in room air, which then mixes with the oxygen. 
The larger the size of holes in the Venturi, the more room air is 
drawn in. The resulting mixture then flows into the Smart Oxygen 
unit providing the animal with an oxygen rich environment.  Each 
Venturi is labeled with the percent of oxygen it can provide.

INSTRUCTIONS FOR OXYGEN THERAPY

u

  The Smart Oxygen unit has two ports 

A

 that may be used as inlet 

and outlet. Attach the double male connector 

B

 to one port 

A

.

u

  Connect the tubing to your oxygen source.  

 

u

  If using an anesthesia machine - Connect the blank    

 

  Venturi (no holes) 

C

 to one end of the tubing 

D

 and attach    

  to the end of the anesthesia breathing circuit.

 

u

  If using a regulator/flowmeter with a male DISS 

    outlet – Connect the Christmas Tree Swivel Adaptor 

G

 to  

 

  one end of the tubing 

D

 and attach to your oxygen supply.

 

u

  If using a regulator/flowmeter with a barbed outlet –     

 

  Connect one end of the tubing 

D

 directly to your oxygen  

 

 supply.

Содержание J1509

Страница 1: ...SMART OXYGEN CAGE J1509 USER MANUAL Oxygen Therapy An important modality for critically ill patients...

Страница 2: ...gen therapy and nebulization Figure 1 SMART OXYGEN OVERVIEW CONTENTS A I J A B C D E G F u A u B u C u D u E u F u G u H u I u J 2x Elbow Outlets w Cups Double Male Connector Venturi Oxygen Blank Opaq...

Страница 3: ...outlet Connect one end of the tubing D directly to your oxygen supply u Connect the green Venturi E to the other end of the tubing and attach it to the Smart Oxygen inlet port via the double male conn...

Страница 4: ...enturi As the oxygen passes though the Venturi it draws in room air which then mixes with the oxygen The larger the size of holes in the Venturi the more room air is drawn in The resulting mixture the...

Страница 5: ...the oxygen at the required flow rate Ensure the outlet port is uncapped to prevent pressure and CO2 from building up in the system Attach waste gas anesthesia scavenging tubing if necessary u Allow t...

Страница 6: ...ttach the double male connector B to one port and the other to your waste gas anesthesia scavenging tubing u Connect the tubing to your oxygen source u If using an anesthesia machine Connect the blank...

Страница 7: ...sh the unit with oxygen before lifting the lid Failure to do so may result in medication or gas escaping into the procedure area Patient breathing rate should be monitored as it may change due to anxi...

Страница 8: ...determine the minimum oxygen flow rate needed to maintain the oxygen concentration base on the patient s current respiratory rate u Example 2 4 lpm minimum oxygen flow rate 1 min to maintain oxygen u...

Страница 9: ...f the display unit to select Celsius or Fahrenheit Humidity is displayed in percent The lowest and highest temperature humidity recorded over past 24 hours is also displayed along the right side of th...

Страница 10: ......

Страница 11: ......

Страница 12: ...Jorgensen Laboratories Inc Loveland CO 80538 800 525 5614 fax 970 663 5042 www jorvet com info jorvet com Professional Veterinary Equipment Supplier...

Отзывы: