Unscrewing the handpiece from the micromotor
The handpiece is screwed onto the mircromotor. A hook spanner is used for loosening.
Figure 7.4 Schematic drawing
Quick coupling system joke Rapid-Change for ENESKA 3-2!
• just pull out the old handpiece an push in the new one
• modify the motor and handpiece in two steps only:
connector plug und coupling (order no. 0 007 051 and 0 007 050).
Figure 7.5 Rapid-Change
loosen
screw tight
micromotor
handpiece
Fuse replacement
Pull out the mains cable (7.2/3) from the socket.
The 230 V mains fuse is located behind the cover (7.2/2). Pull out the fuse holder on the rear of the
unit.
Replace the fuse. The fuse must comply with specifi cation 250 Vac, T1A.