background image

9 - 11 

MAINTENANCE PROCEDURE G260 – REVISION 02 - 2013-04 

 

ADJUSTMENT 

 

WARNING: To adjust a 2

nd

 stage correctly, it’s must be connected to a appropriate 

and well adjusted 1

st

 stage (with a interstage pressure between 9,3 - 9,7 bar) and a 

tank with a minimal filling pressure of 200 bar. 

 
1. 

Check that the adjustment knob is still screwed out completely to achieve the lightest 
breathing resistance. The position the V.I.V.A. flow vane has no importance. 

 
2. 

Before connecting the 2

nd

 stage adjustment tool to the 2

nd

 stage and without pushing 

on the purge, slowly blow through the inlet tube to detect a leak. This information will 
indicate or the orifice is in contact (no leak) or not (leak) with the seat. 
 
If a leak is detected, screw the 2

nd

 stage adjustment tool to the 2

nd

 stage and slowly 

screw the orifice one turn clockwise and repeat the test. If there is no leakage, the 2

nd

 

stage can be adjusted. 

 
3. 

Connect the Digi-Mag adapter mouthpiece to the G260 mouthpiece and turn on the 
Digi-Mag. Zeroize the Digi-Mag pressure sensor if needed. 

 
4. 

Connect the appropriate and well adjusted 1

st

 stage to a tank with a minimal filling 

pressure of 200 bar, connect the 1

st

 stage to the 2

nd

 stage adjustment tool with the 

L.P. hose and 

slowly

 open the tank valve. 

 
5. Push 

slowly

 on the tool adjustment knob and turn slowly the knob to look for the 

groove of the orifice. 
 
As soon as the groove is found, slowly unscrew the adjustable orifice - counter 
clockwise - until a small leak is detected, then screw in the orifice - just enough 
clockwise - to stop the leak. 

 
6. 

Cycle the 2

nd

 stage 10 times or more by pushing on the purge, then readjust as indi-

cated above. 

 
7. 

Inhale slowly through the Digi-Mag and note the cracking effort. It should be between 
1,0 to 1,2 inches of water (2,5 to 3,0 milibars). 

 

8. 

If the cracking effort is higher then indicated, use the 4 
mm Allen key and unscrew by a ¼ turn the fine adjust-
ment screw inside the adjustment knob. 
 

note :

 The adjustable orifice 

MUST

 always be readjusted as per paragraph 5 and 6 

before measuring the cracking effort again. 
 
Proceed by successive steps to reach the recommended values(1,0 to 1,2 inches of 
water or 2,5 to 3,0 milibars). 

 

Содержание SCUBAPRO G260

Страница 1: ...MAINTENANCE PROCEDURE G260 revision 02 2013 04 ...

Страница 2: ...2 11 MAINTENANCE PROCEDURE G260 REVISION 02 2013 04 ...

Страница 3: ...l 41 043 010 Grease Christo Lube MCG111 41 493 125 Screwdriver ball end 3 32 43 002 101 Universal tool regulator new model 43 040 000 G260 cover tool 43 300 020C O ring extractor 43 300 107 VIVA flow vane extractor 43 300 225 Regulator hose tool on 2nd stage 43 300 235 Counter mat 47 010 000 Digi Mag 6000 Digital Magnehelic SCUBAPRO Magnehelic mouthpiece 43 404 001P Allen key 4 mm Cutter small mod...

Страница 4: ...the adjustment knob 14 Remove with the small screwdriver the plug from the adjustment knob 15 Unscrew with the 4 mm Allen key the fine adjustment screw from the adjustment knob and remove the o ring from the fine adjustment screw 16 Push carefully with the small screwdriver through the orifice to remove the poppet assy spring and balance chamber out of the inlet tube and remove the seat and 2 o ri...

Страница 5: ...he tool against the shoulder of the flow vane the angled portion of the tool should rest on the mouthpiece tube and use the tool as a lever to push the flow vane out of the housing Remove the o ring from the flow vane 22 To remove the exhaust cover use the ball end screwdriver to push the pin out of the exhaust cover and case 23 It is not necessary to carry out this step unless the exhaust valve n...

Страница 6: ...y on the lip at several places and make sure the outer part of the valve rests on the regulator case 2 Place the exhaust cover and insert the pin making sure it is correctly centred 3 If the flow vane has been removed it must be replaced Place a new o ring Do not lubricate the o ring in order not to have a loose flow vane Make sure that the flow vane is right in front of the cavity s in the case a...

Страница 7: ...oulder configuration is required 2 actions must be taken a Rotate the flow switch to open the injection hole facing the technician when the treaded side of the inlet tube is on the left b Place the adapter on the left side of the case For the standard RIGHT shoulder configuration c Rotate the flow switch to open the injection hole facing the technician when the treaded side of the inlet tube is on...

Страница 8: ... inside the case placing the lever facing the diaphragm Check if the injection hole is facing the mouthpiece tube 19 Slightly lubricate the inlet tube threads Screw the nut and tighten moderately with the regulator hose tool 20 Place the o ring over the thread of the inlet tube 21 Position the diaphragm making sure that it is not distorted 22 Assemble the spring and purge button in the cover assem...

Страница 9: ...i Mag Zeroize the Digi Mag pressure sensor if needed 4 Connect the appropriate and well adjusted 1st stage to a tank with a minimal filling pressure of 200 bar connect the 1st stage to the 2nd stage adjustment tool with the L P hose and slowly open the tank valve 5 Push slowly on the tool adjustment knob and turn slowly the knob to look for the groove of the orifice As soon as the groove is found ...

Страница 10: ...e to push the purge button 1 mm with out a leakage of the 2nd stage 11 Remove the adapter mouthpiece of the Digi Mag from the G260 mouthpiece 12 Fit the plug in the adjustment knob 13 Close the tank valve purge and remove the 2nd stage adjustment tool note before removing the 2nd stage adjustment tool make sure that the adjustment knob is retracted out of the groove in the orifice to ensure that t...

Страница 11: ...11 11 MAINTENANCE PROCEDURE G260 REVISION 02 2013 04 ...

Отзывы: