Titanium Cannulated Tibial Nail Expert Nailing System Surgical Technique DePuy Synthes Trauma 39
Proximal locking— Proximal Segment fractures
5
Insert dual core locking screws
Instrument
03.010.107
Stardrive Screwdriver, t25, self-retaining
Insert the appropriate 5.0 mm dual core locking screw
through the protection sleeve,
using the t25 Stardrive
screwdriver. do not overtighten.
verify locking screw length under image intensification.
repeat this procedure for the second dual core locking screw.