18 DePuy Synthes MATRIX Spine System – Degenerative Surgical Technique
Optional Technique
Adding Rod-to-Rod Connectors
Notes:
•
If any part of the construct requires further ad-
justment, all set screws must be loosened to the
point of resistance. Do not remove set screws from
the assembly. After final adjustment, retighten the
set screws.
•
Parallel connectors contain Nitinol components.
Implants that contain Nitinol should not be used
in patients with nickel sensitivities or allergies.
•
The retaining sleeve for transverse connectors
cannot be used when tightening parallel with two
set screws.
Precautions:
•
Care should be taken not to tighten the connector
on a portion of the rod that has been contoured or
deformed by a rod cutter.
•
Synthes connectors are tested using Synthes 5.5 or
6.0 mm rods and not compatible with those rods of
similar diameter from other manufacturers.
•
Refer to the torque-limiting handle package and
labeling for the recommended calibration mainte-
nance.
Optional instrument
03.632.052
Screwdriver Stardrive, T15, short,
for Matrix