11 DePuy Synthes MATRIX Spine System – Degenerative Surgical Technique
Surgical Technique
10. Compress
Instruments
388.422
Compression Forceps, length 335 mm,
for Pedicle Screws
03.632.401
Screwdriver Shaft Stardrive, T25, long,
straight tip, with Hexagonal Coupling,
for Matrix
03.620.061
T-Handle with Ratchet Wrench
and with Torque Limiter, 10 Nm
03.632.049
Counter Torque, standard,
for Matrix 5.5
Precaution:
Ensure all locking caps are fully re-
duced and provisionally tightened (see “Insert
1-step locking cap”). Failure to do so could poten-
tially lead to a misalignment.
Finally tighten one locking cap completely to create a
fixed point for distraction (see “Perform final tighten
-
ing”). Reverse the locking cap of the screw to be relo-
cated a quarter of a turn.
Use the compression forceps to compress the construct.
Once in the desired position, tighten the locking caps
with the screwdriver.
The rod holding forceps can be used as a temporary
point of compression when adjacent pedicle screws are
too distant.
Note:
Always fully seat the counter torque on the
rod. The instrument must be perpendicular to the
rod during tightening.
Optional instrument
03.632.202
Holding Forceps for Rods
B
5.5
and
B
6.0 mm
Correct
Incorrect