16 DePuy Synthes INSIGHT Lateral Access System Surgical Technique
Retraction
Attach Connector
Instruments
03.816.801
Connector, for Strong Arm
No. 03.816.800
03.816.010
Screwdriver, for INSIGHT Lateral Access
System
Optional
03.816.019
Wrench, for INSIGHT Lateral Access
System
Attach connector to the respective point on the retractor
(A or B, see next page) and tighten with the Screwdriver
or wrench depending on tightening preference (1). The
connector can only be attached in the illustrated posi-
tions and orientations (from the left and right side) to
avoid subsequent conflict with angulation mechanism.
For the role of the different attachment options, see
next page.