Hoyer
®
Switch
3
Joerns Healthcare Ltd recommends the Hoyer Switch and its accessories are subjected to inspection and
maintenance, as detailed in this manual, at appropriate intervals depending on a risk assessment of the
frequency of use and environmental factors. This will ensure the equipment is adequately maintained and
remains safe for use as detailed in PUWER (Provision and Use of Work Equipment Regulations). The risk
assessment is necessary as PUWER does not detail specific time intervals.
CAUTION
In the event of identifying any faulty items, please contact your local authorized service provider
or Joerns Healthcare directly.
Frame
• Check the operation of the sliding section quick release pins.
• Make sure the frame is fully engaged into the base supports.
• Ensure both frame locking knobs are fully tightened.
Casters and Wheels
•
Check all casters and wheels for firm attachment to the base.
• Check for free rotation of the wheels and the casters swivel.
• Check for excessive wear, cracking or splits on the center wheel outer tread and the corner caster
wheels.
•
Where possible, remove any build up of threads, hair or fibers.
• Lubricate the swivel and axle bearings, if necessary, with a light mineral based grease or food grade
spray lubricant.
• Check correct operation of the brakes.
Cleaning
Clean with ordinary soap and water and/or any hard surface disinfectant. Harsh chemical cleaners or
abrasives should be avoided as these may damage the surface finish of the product. After cleaning, the
unit should be thoroughly dried.
WARNING
HOYER RECOMMENDS THE USE OF GENUINE HOYER PARTS. Hoyer sling/belt and products are
designed to be compatible with one another. For country specific guidance on sling/belt use and
compatibility, please refer to the sling/belt label or contact your local market distributor or Joerns
Healthcare.
Inspection Criteria