Guidance and Manufacturer’s Declaration - Electromagnetic Immunity:

The AutoVector

®

 Fluid Immersion Simulator is intended for use in the electromagnetic environment specified below. 

The customer or the user of the Auto Vector Fluid Immersion Simulator should assure that it is used in such an 

environment.

Immunity Test

IEC60601 Test Level Compliance

Electromagnetic Environment-Guidance

Electrostatic 

discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete or ceramic tile. 

If floors are covered with synthetic material, the 

relative humidity should be at least 30%. 

Electrical fast 

transient/burst 

IEC 61000-4-4

±2 kV for power 

supply lines

±1 kV for input/output 

lines

±2 kV for power 

supply lines

Not applicable

Not applicable  Mains power quality should 

be that of a typical commercial or hospital 

environment.

Surge 

IEC 61000-4-5

±1 kV line(s) to 

line(s)

±2 kV line(s) to earth

±1 kV line(s) to 

line(s)

±2 kV line(s) to earth

Mains power quality should be that of a typical 

commercial or hospital environment.

Voltage 

dips, short 

interruptions 

and voltage 

variations on 

power supply 

input lines IEC 

61000-4-11

<5% 

U

τ 

(>95% dip in 

U

τ) 

for 0,5 cycle 

40% 

U

τ 

(60% dip in 

U

τ) 

for 5 cycles 

70% 

U

τ 

(30% dip in 

U

τ) 

for 25 cycles 

<5% 

U

τ

(>95% dip in 

U

τ) 

for 5 sec 

<5% 

U

τ 

(>95% dip in 

U

τ) 

for 0,5 cycle 

40% 

U

τ 

(60% dip in 

U

τ) 

for 5 cycles 

70% 

U

τ 

(30% dip in 

U

τ) 

for 25 cycles 

<5% 

U

τ 

(>95% dip in 

U

τ) 

for 5 sec 

Mains power quality should be that of a typical 

commercial or hospital environment. If the user 

of the AutoVector Fluid Immersion Simulator 

requires continued operation during power mains 

interruptions, it is recommended that the Auto 

Vector Fluid Immersion Simulator be powered 

from an uninterruptible power supply or a battery. 

Power frequency 

(50/60 Hz)  

magnetic field 

IEC 61000-4-8

3 A / m

Not applicable

Power frequency magnetic fields should be at 

levels characteristic of a typical location in a 

typical commercial or hospital environment.

Note:

 

U

τ is the a.c. mains voltage prior to application of the test level.

Содержание Dolphin Fluid Immersion Simulation System

Страница 1: ...dvanced Surfaces Dolphin Fluid Immersion Simulation System Dolphin Pediatric Fluid Immersion Simulation System To avoid injury read user s manual before using Representative image of the Dolphin Fluid...

Страница 2: ...y within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used Danger Explosion Hazard Do not use in the presence of flamma...

Страница 3: ...for assessment and implementation of bed side rails in various health care settings Entrapment zones involve the relationship of bed components often directly assembled by the healthcare facility rath...

Страница 4: ...pecialty Surface Wheelchair Cushion Troubleshooting 12 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures 12 13 Recommended Linen Changing the The...

Страница 5: ...on FIS System is an advanced therapy system designed to provide state of the art pressure redistribution by simulating the effects of a body immersed in a fluid medium The Dolphin System includes thre...

Страница 6: ...bution technology designed to alleviate vertical shear forces The therapy mattresses and specialty surfaces conform to the specific shape of the patient minimizing soft tissue distortion reducing isch...

Страница 7: ...ess that can fit any healthcare bed frame including bariatric frames up to 48 122 cm wide Conforms to specific shape of the patient minimizing soft tissue distortion reducing ischial tuberosity penetr...

Страница 8: ...battery back up will provide power for continued therapy mattress system operation for approximately 12 hours Keyboard Functions Warning For important precautions see page two Caution The patient s he...

Страница 9: ...d from further adjustments When locked pressing the Lock button again restores normal operation and the yellow LED is extinguished Battery Indicators Figure 4 The Battery indicator will blink when the...

Страница 10: ...0 LED s and cover the full control range from Soft to Firm When operating within normal parameters the Auto Feedback LED scale will be amber Should the system be outside of normal parameters the LED s...

Страница 11: ...e appropriate position based on the patient s clinical condition 2 Select the highest or most firm Comfort Adjust setting 3 Hand Check Place a hand with three 3 fingers if head of bed at 30 or higher...

Страница 12: ...g on Battery Back up Ensure that the power is not on Standby If on Standby press the Power button Ensure that all air cells are connected Ensure that the Mode is in the appropriate position for the at...

Страница 13: ...that the patient is positioned properly within the confines of the bed or other specialty item Do not let any extremities protrude over the side or between the bed rails when the mattress is being use...

Страница 14: ...he therapy mattress or specialty surface deflated thoroughly wipe down entire mattress as air cells will lie flat Be sure to reach all areas underneath and in between air cells Allow to air dry If dus...

Страница 15: ...retcher Pad Maximum weight capacity2 700 lbs 318 1 Kg Wheelchair Cushion Maximum weight capacity3 250 lbs 113 6 Kg Pediatric System Minimum weight capacity1 5 lbs 2 3 Kg Dimensions Control Unit 11 5 2...

Страница 16: ...should be observed to verify normal operation in the configuration in which it will be used Guidance and Manufacturer s Declaration Electromagnetic Emissions The AutoVector Fluid Immersion Simulator...

Страница 17: ...ospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital env...

Страница 18: ...ation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference...

Страница 19: ...ng to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz...

Страница 20: ...e Improper maintenance handling and or cleaning Failure to use in the manner indicated in the Dolphin FIS System user manual Any modification repair or alteration done to the Dolphin FIS System that w...

Отзывы: