Joerns® Advanced Support Surfaces
Dolphin Fluid Immersion Simulation® Systems
© 2014 Joerns Healthcare • 6110109 RevG • 14-2640
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity:
The AutoVector
®
Fluid Immersion Simulator is intended for use in the electromagnetic environment specified below.
The customer or the user of the Auto Vector Fluid Immersion Simulator should assure that it is used in such an
environment.
Immunity Test
IEC60601 Test Level Compliance
Electromagnetic Environment-Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
Not applicable
Not applicable Mains power quality should
be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
<5%
U
τ
(>95% dip in
U
τ)
for 0,5 cycle
40%
U
τ
(60% dip in
U
τ)
for 5 cycles
70%
U
τ
(30% dip in
U
τ)
for 25 cycles
<5%
U
τ
(>95% dip in
U
τ)
for 5 sec
<5%
U
τ
(>95% dip in
U
τ)
for 0,5 cycle
40%
U
τ
(60% dip in
U
τ)
for 5 cycles
70%
U
τ
(30% dip in
U
τ)
for 25 cycles
<5%
U
τ
(>95% dip in
U
τ)
for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the AutoVector Fluid Immersion Simulator
requires continued operation during power mains
interruptions, it is recommended that the Auto
Vector Fluid Immersion Simulator be powered
from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A / m
Not applicable
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Note:
U
τ is the a.c. mains voltage prior to application of the test level.