Appendix A: Electromagnetic Compatibility (EMC) Information

AutoVector

®

 Fluid Immersion Simulator

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service 

according to the EMC information provided in this manual.

 

Portable and mobile RF communications equipment can affect Medical Electrical Equipment.

The use of Accessories, transducers, and cables other than those specified by the manufacturer, may result in 

increased Emissions or decreased Immunity of the Auto Vector Fluid Immersion Simulator.

The AutoVector Fluid Immersion Simulator should not be used adjacent to or stacked with other equipment and that if 

adjacent or stacked use is necessary, the Auto Vector Fluid Immersion Simulator should be observed to verify normal 

operation in the configuration in which it will be used.

Guidance and Manufacturer’s Declaration –Electromagnetic Emissions

The AutoVector Fluid Immersion Simulator is intended for use in the electromagnetic environment specified below. 

The customer or the user of the Auto Vector Fluid Immersion Simulator should assure that it is used in such an 

environment. 

Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions 

CISPR 11

Group 1

The AutoVector Fluid Immersion Simulator uses RF energy only for its 

internal function. Therefore, its RF emissions are very low and are not 

likely to cause any interference in nearby electronic equipment.

RF emissions 

CISPR 11

Class B

The AutoVector Fluid Immersion Simulator is suitable for use in all 

establishments, including domestic establishments and those directly 

connected to the public low-voltage power supply network that supplies 

buildings used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations / 

flicker emissions 

IEC 61000-3-3

Complies

Содержание Dolphin Fluid Immersion Simulation System

Страница 1: ...dvanced Surfaces Dolphin Fluid Immersion Simulation System Dolphin Pediatric Fluid Immersion Simulation System To avoid injury read user s manual before using Representative image of the Dolphin Fluid...

Страница 2: ...y within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used Danger Explosion Hazard Do not use in the presence of flamma...

Страница 3: ...for assessment and implementation of bed side rails in various health care settings Entrapment zones involve the relationship of bed components often directly assembled by the healthcare facility rath...

Страница 4: ...pecialty Surface Wheelchair Cushion Troubleshooting 12 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures 12 13 Recommended Linen Changing the The...

Страница 5: ...on FIS System is an advanced therapy system designed to provide state of the art pressure redistribution by simulating the effects of a body immersed in a fluid medium The Dolphin System includes thre...

Страница 6: ...bution technology designed to alleviate vertical shear forces The therapy mattresses and specialty surfaces conform to the specific shape of the patient minimizing soft tissue distortion reducing isch...

Страница 7: ...ess that can fit any healthcare bed frame including bariatric frames up to 48 122 cm wide Conforms to specific shape of the patient minimizing soft tissue distortion reducing ischial tuberosity penetr...

Страница 8: ...battery back up will provide power for continued therapy mattress system operation for approximately 12 hours Keyboard Functions Warning For important precautions see page two Caution The patient s he...

Страница 9: ...d from further adjustments When locked pressing the Lock button again restores normal operation and the yellow LED is extinguished Battery Indicators Figure 4 The Battery indicator will blink when the...

Страница 10: ...0 LED s and cover the full control range from Soft to Firm When operating within normal parameters the Auto Feedback LED scale will be amber Should the system be outside of normal parameters the LED s...

Страница 11: ...e appropriate position based on the patient s clinical condition 2 Select the highest or most firm Comfort Adjust setting 3 Hand Check Place a hand with three 3 fingers if head of bed at 30 or higher...

Страница 12: ...g on Battery Back up Ensure that the power is not on Standby If on Standby press the Power button Ensure that all air cells are connected Ensure that the Mode is in the appropriate position for the at...

Страница 13: ...that the patient is positioned properly within the confines of the bed or other specialty item Do not let any extremities protrude over the side or between the bed rails when the mattress is being use...

Страница 14: ...he therapy mattress or specialty surface deflated thoroughly wipe down entire mattress as air cells will lie flat Be sure to reach all areas underneath and in between air cells Allow to air dry If dus...

Страница 15: ...retcher Pad Maximum weight capacity2 700 lbs 318 1 Kg Wheelchair Cushion Maximum weight capacity3 250 lbs 113 6 Kg Pediatric System Minimum weight capacity1 5 lbs 2 3 Kg Dimensions Control Unit 11 5 2...

Страница 16: ...should be observed to verify normal operation in the configuration in which it will be used Guidance and Manufacturer s Declaration Electromagnetic Emissions The AutoVector Fluid Immersion Simulator...

Страница 17: ...ospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital env...

Страница 18: ...ation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference...

Страница 19: ...ng to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz...

Страница 20: ...e Improper maintenance handling and or cleaning Failure to use in the manner indicated in the Dolphin FIS System user manual Any modification repair or alteration done to the Dolphin FIS System that w...